Monitor unit calculation for tangential breast treatments: Verification in an anthropomorphic phantom

نویسندگان

  • Stephen J. Howlett
  • Tomas Kron
چکیده

This paper presents an anthropomorphic phantom study of dose delivered to a specific point during tangential breast irradiation to verify monitor unit calculations. Measurements were made using a 0.6 cc Farmer type cylindrical ionization chamber in the phantom and compared to calculations made on a three-dimensional radiotherapy treatment planning system using single digitized contour through to multi slice CT data. A large breast phantom was used for a single field size with a combination of open and wedged fields for three different energies (4, 6, and 18 MV). Solid flat phantom measurements were also performed for comparison. Results showed a lower calculated dose than the dose measured for a fixed number of monitor units where the variations were within a range of 0.8% to 4.5%. Differences were larger for the anthropomorphic phantom than the flat phantom. We conclude that little accuracy is gained from CT based monitor unit calculations compared to those based on digitised contours for this breast treatment but that the dose distributions will be affected. This type of test is recommended as one of a large set, in the commissioning and testing procedures for treatment planning systems.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of treatment planning system monitor unit calculations for three intensity modulated radiotherapy delivery techniques

Background: We have validated the monitor unit calculations from a commercially available treatment planning system (TPS) for three intensity modulated radiotherapy (IMRT) planning techniques for tangential breast irradiation by using ionization chamber measurements. Materials and Methods: Treatment plans were generated for forty-two breast patients by a forward planned field in field ...

متن کامل

Verification of the accuracy of 3D calculations of breast dose during tangential irradiation: measurements in a breast phantom

This report specifically describes the use of a unique anthropomorphic breast phantom to validate the accuracy of three-dimensional dose calculations performed by a commercial treatment-planning system for intact-breast tangential irradiation. The accuracy of monitor-unit calculations has been corroborated using ionization chamber measurements made in this phantom. Measured doses have been comp...

متن کامل

An Approach in Radiation Therapy Treatment Planning: A Fast, GPU-Based Monte Carlo Method

Introduction: An accurate and fast radiation dose calculation is essential for successful radiation radiotherapy. The aim of this study was to implement a new graphic processing unit (GPU) based radiation therapy treatment planning for accurate and fast dose calculation in radiotherapy centers. Materials and Methods: A program was written for parallel runnin...

متن کامل

Verifying monitor unit calculations for tangential breast fields

An essential component of quality assurance in radiation therapy is verifying the accuracy of monitor unit calculations. Differences between between sophisticated algorithms using 2.5D or 3D calculations and simpler Monitor unit check algorithms assuming a flat water phantom must be expected. For many anatomical sites such differences are small and of little or no consequence in the context of ...

متن کامل

Verification of tangential breast treatment dose calculations in a commercial 3D treatment planning system

The accuracy of the photon convolution/superposition dose algorithm employed in a commercial radiation treatment planning system was evaluated for conditions simulating tangential breast treatment. A breast phantom was fabricated from machineable wax and placed on the chest wall of an anthropomorphic phantom. Radiographic film was used to measure the dose distribution at the axial midplane of t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 3  شماره 

صفحات  -

تاریخ انتشار 2002