Retrospective analysis of reasons for failure of DDD pacemaker implantation in patients operated on between 1993 and 2005.
نویسندگان
چکیده
BACKGROUND During implantation of a DDD pacemaker the following difficulties may be encountered: venous anomalies (the absence of vessels of adequate calibre or difficulty in subclavian vein puncture), arrhythmias during implantation (episodes of atrial flutter/fibrillation while the atrial leads are being positioned), lack of mechanical stability of the electrode in the heart chamber and inability to achieve an acceptable pacing and sensing threshold during implantation. The purpose of the study was to analyse retrospectively the reasons for DDD pacemaker failure in patients operated on between 1993 and 2005. METHODS We reviewed retrospectively all implantation data from 1988 to 2005 to identify patients with primary failure of DDD pacemaker implantation. Further analysis included patients who had received a DDD pacemaker between 1993 and 2005, when this type of pacemaker made up between 9 and 40% of all pacemaker implantations. We implanted 7469 pacemakers, including 1958 (26.2%) dual-chamber pacemakers, in 783 patients with atrioventricular block (AVB), 392 with sick sinus syndrome (SSS), 450 with AVB +/- SSS and 333 with tachy-brady syndrome (TBS). The mean age of the patients was 65.5 +/- 17.3 years. DDD pacing was unsuccessful in 108 (1.4%) patients, including 32 with AVB, 22 with SSS, 16 with SSS +/- AVB and 38 with TBS. The mean age of these patients was 78.5 +/- 19.4 years. RESULTS The reasons for failed implantation were venous anomalies in 12%, an arrhythmia episode in 27.8%, a high pacing threshold in the atrium in 17.6%, low atrial potential amplitude in 25.9% and lack of mechanical stability of the electrode in 16.7% of patients. The difficulties were encountered in elderly patients (p < 0.01), most frequently in patients with SSS and TBS (71). Between 2004 and 2005 venous anomalies and a high pacing threshold were the main causes of failure. CONCLUSIONS Currently the main difficulties encountered during pacemaker implantation are venous anomalies and a high pacing threshold. Arrhythmia episodes, low atrial potential amplitude and lack of mechanical stability are of minor importance. Elderly patients with sick sinus syndrome and tachy-brady syndrome have the highest failure rate. (Cardiol J 2007; 14: 155-159).
منابع مشابه
CARDIOVASCULAR MEDICINE Long term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity
Objective: To determine how short and long term complication rates after pacemaker implantation are influenced by patient morbidity, operator experience, and choice of pacing system. Design: Retrospective analysis of 1884 patients who received VVI (n = 610), VDD (n = 371), or DDD devices (n = 903) between 1990 and 2001. Follow up period was 64 (34) months. The influence of age, sex, coronary ar...
متن کاملComparative Study of Electrical Stimulation of the Heart with VDD and DDD Pacemakers as to the Evolution to Atrial Fibrillation
INTRODUCTION The pacemaker implantation VDD is considered simpler, faster, less expensive and causes fewer complications compared to DDD. However, the VDD pacemaker has not been widely used in many centers, perhaps for fear of dysfunction of the sinus node and the reduction of atrial sensitivity by the pacemaker during follow-up after implantation. OBJECTIVE To compare patients with DDD and V...
متن کاملبررسی مشکلات جسمی، روانی و اجتماعی بیماران دارای دستگاه ضربا ساز قلب
The aim of this cross sectional study is the assessment of physical and psychosocial problems of patients with pacemaker and effects of social - individual factors (such as age, sex, marital status, level of education,....) on the intensity of problems. In the study 60 patients (27 women, 33 men) that their mean age was 60.07 years, were studied. Sampling was purposive and the samples were se...
متن کاملLong term complications in single and dual chamber pacing are influenced by surgical experience and patient morbidity.
OBJECTIVE To determine how short and long term complication rates after pacemaker implantation are influenced by patient morbidity, operator experience, and choice of pacing system. DESIGN Retrospective analysis of 1884 patients who received VVI (n = 610), VDD (n = 371), or DDD devices (n = 903) between 1990 and 2001. Follow up period was 64 (34) months. The influence of age, sex, coronary ar...
متن کاملTrends in permanent pacemaker implantation in the United States from 1993 to 2009: increasing complexity of patients and procedures.
OBJECTIVES This study sought to define contemporary trends in permanent pacemaker use by analyzing a large national database. BACKGROUND The Medicare National Coverage Determination for permanent pacemaker, which emphasized single-chamber pacing, has not changed significantly since 1985. We sought to define contemporary trends in permanent pacemaker use by analyzing a large national database....
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Cardiology journal
دوره 14 2 شماره
صفحات -
تاریخ انتشار 2007