Fatal septicemia in a patient with cerebral lymphoma and an Amplatzer septal occluder: a case report
نویسندگان
چکیده
INTRODUCTION The Amplatzer septal occluder is frequently used for percutaneous closure of an atrial septal defect. Complications include thrombosis and embolism, dislocation, cardiac perforation, and, rarely, infection. We report the case of a patient who had survived an occluder-related thromboembolism two years previously. CASE PRESENTATION A 72-year-old Caucasian woman had received a septal occluder because of an atrial septal defect seven years ago. Two years ago, she underwent chemotherapy of a non-Hodgkin lymphoma, developed atrial fibrillation, and experienced a left-sided occluder thrombosis with stroke and peripheral embolism. Now, she presented with cerebral lymphoma, received glucocorticoids, and subsequently developed skin lesions. Swabs from the lesions and blood cultures were positive for methicillin-resistant Staphylococcus aureus and Pseudomonas aeruginosa. Endocarditis, however, was considered only two months later and echocardiography suggested aortic valve endocarditis. Despite antibiotic therapy, she died three days later because of septicemia, and no post-mortem investigation was carried out. It remains uncertain whether the septal occluder was endothelialized or infected and whether explantation might have changed the outcome. CONCLUSIONS If infections occur in patients with a septal occluder, endocarditis should be considered and echocardiography should be performed early. To prevent a fatal outcome, explantation of the septal occluder should be considered, especially in patients with problems that suggest delayed endothelialization. Post-mortem investigations, including bacteriologic studies, should be carried out in patients with a septal occluder in order to assess the focal and global long-term effects of these devices.
منابع مشابه
[Simultaneous percutaneous closure of patent foramen ovale and left atrial appendage].
We report the case of a 59-year-old male smoker with hypertension and dyslipidemia who, in 2007, was diagnosed with atrial fibrillation and patent foramen ovale (PFO) after suffering a cerebral ischemic attack in the region of the middle cerebral artery. The high risk of a further cerebral ischemic attack (score of 3 on the CHADS 2) was an indication for long-term oral anticoagulation. However,...
متن کاملPenetration of left and right atrial wall and aortic root by an Amplatzer atrial septal occluder in a nine year old boy with Marfan syndrome: Case report
BACKGROUND To describe complications associated with Amplatzer septal occluders in a patient with Marfan syndrome CASE PRESENTATION A nine-year-old boy with Marfan syndrome and a 22 mm atrial septal defect (ASD) was treated successfully by interventional closure of his ASD by placing a 24 mm Amplatzer septal occluder. Follow up examinations showed a good result but an increasing enlargement o...
متن کاملTranscatheter closure of post-myocardial infarction ventricular septal defect with Amplatzer septal occluder.
Rupture of the interventricular septum is an uncommon but often fatal complication of acute myocardial infarction. Transcatheter closure is an established method of treating selected congenital defects but clinical experience on transcatheter closure of postinfarction ventricular septal defects (VSDs) is minimal. We report a case of successful transcatheter closure of postinfarction VSD using t...
متن کاملAmplatzer septal occluder-induced transient complete atrioventricular block.
Percutaneous transcatheter atrial septal defect (ASD) closure is a widely used technique that has replaced open-heart surgical closure in many centers. The most common implant is the Amplatzer septal occluder which seems to be a highly effective and safe device. However, there are reports of complications associated with its implantation. We report a 9-year-old boy who presented with complete a...
متن کاملPercutaneous Retrieval of Embolized Amplatzer Septal Occluder after Treatment of Double Atrial Septal Defect: A Case Report
Embolization of the occlusion device after percutaneous closure of atrial septal defect (ASD) is a potential disastrous complication. The usual site of embolization is the right side of the heart including pulmonary artery, but the device embolization to the extracardiac aorta is extremely rare. Here, we report a successful percutaneous retrieval case of the embolized Amplatzer Septal Occluder ...
متن کامل