Medtronic CapSureFix MRI 5086 pacing lead (5086; Medtronic, Inc., Minneapolis, MN, USA) has been reported to be associated with increased cardiac perforation and lead dislodgement. This study aimed to compare the incidence of cardiac perforation and lead dislodgement within 30 days after pacemaker implantation between 5086 MRI lead and previous Medtronic CapSureFix Novus 5076 non-MRI pacing lead. This was a nationwide, multicenter retrospective study in which we compared the incidence of adverse events between 277 patients implanted with 5086 lead and 205 patients implanted with 5076 lead between March 2009 and September 2014. Cardiac perforation within 30 days of pacemaker implantation occurred in 4 patients (1.4%) with the 5086 lead and in no patient with the 5076 lead (P = 0.084). Lead dislodgement occurred in 8 patients (2.9%) with the 5086 lead and in 5 patients (2.4%) with the 5076 lead (P = 0.764). On multivariate logistic regression analysis, age was significantly associated with cardiac perforation. Congestive heart failure and implantation of right atrial (RA) lead at RA free wall or septum were significant factors for the incidence of lead dislodgement and lead revision. The incidence of cardiac perforation and lead dislodgement were not statistically different between the patients with 5086 lead and the patients with 5076 lead. However, careful attention for cardiac perforation may be needed when using the 5086 MRI lead, especially in elderly patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pacemaker lead-straight through the heart.

1 of 2 DESCRIPTION A 45-year-old female patient, was admitted with recurrent syncope of 1 month duration. An emergency temporary transvenous right ventricular pacing electrode was placed for complete heart block. The patient developed chest pain and giddiness the following day and electrocardiogram revealed failure to capture and undersensing, with right bundle branch block morphology ( fi gure...

متن کامل

Incidence of lead dislodgement, malfunction and perforation during the first year following device implantation

BACKGROUND The number of cardiac rhythm device implantations has been growing fast due to expanding indications and ageing of the population. Complications of implantation were rare in the trials. However, these involved small numbers and selected patients. Prospective real-life data are necessary to assess cardiac device implantation procedure-related risks. OBJECTIVE To determine the incide...

متن کامل

Leadless cardiac pacing

Transvenous cardiac pacemakers definitely improve quality of life and reduce mortality in at-risk patients, but they are associated with several potential device-related complications. Approximately 10% of patients experience complications related to transvenous implantation of the pacemaker. These may be attributable to either the pulse generator (hematoma, skin breakdown, pocket infection) or...

متن کامل

Use of a novel pacing mode to achieve biventricular pacing in a patient with recurrent atrial lead dislodgement after CRT-D implantation

Cardiac resynchronization therapy device (CRT-P and CRT-D) implantation has increased tremendously with increasing operator experience, eligible patients and expansion of indications. Refinements in devices and algorithms now aid physicians to improve biventricular pacing and optimize CRT. We report a case in which an interesting device program was used to achieve biventricular pacing after rep...

متن کامل

Use and safety of magnetic resonance imaging in patients with an implanted MRI-conditional pacing device

Methods This was a retrospective, single center study. All patients with an MRI conditional pacemaker or Internal Cardioverter Defibrillator (ICD) implanted in our center between June 2011 and March 2015 were enrolled. All MRI examinations performed in these patients were evaluated. For all patients, pacing thresholds, Pand R-wave amplitude, lead impedance and battery longevity were monitored d...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 31  شماره 

صفحات  -

تاریخ انتشار 2016