Hardness evaluation of prosthetic silicones containing opacifiers following chemical disinfection and accelerated aging.

نویسندگان

  • Marcelo Coelho Goiato
  • Marcela Filié Haddad
  • Daniela Micheline dos Santos
  • Aldiéris Alves Pesqueira
  • Amália Moreno
چکیده

We evaluated the effects of disinfection and aging on the hardness of silicones containing opacifiers and intended for use in facial prosthetics. A total of 90 samples were produced using a cylindrical metal mold 3 mm in height and 30 mm in diameter. The samples were fabricated from Silastic MDX 4-4210 silicone in three groups: GI contained no opacifier, GII contained barium sulfate (Ba), and GIII contained titanium dioxide (Ti). The samples were disinfected using effervescent tablets (Ef), neutral soap (Ns), or 4% chlorhexidine (Cl) 3 times a week for 60 days. After this period the samples underwent 1,008 hours of accelerated aging. The hardness was measured using a durometer immediately following the disinfection period and after 252, 504, and 1,008 hours of aging. The data were statistically analyzed using 3-way ANOVA and the Tukey test (p < .05). The GIII group exhibited the greatest variation in hardness regardless of elapsed time. All groups displayed greater hardness after 1,008 hours of accelerated aging independent of disinfectant type. All of the hardness values were within the clinically acceptable range.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Influence of opacifiers on dimensional stability and detail reproduction of maxillofacial silicone elastomer

BACKGROUND We evaluated the influence of chemical disinfection and accelerated aging on the dimensional stability and detail reproduction of a silicone elastomer containing one of two opacifiers. METHODS A total of 90 samples were fabricated from Silastic MDX 4-4210 silicone and divided into groups (n = 10) according to opacifier content (barium sulfate or titanium dioxide) and disinfectant s...

متن کامل

Evaluation of hardness and surface roughness of two maxillofacial silicones following disinfection.

The purpose of this study was to evaluate the Shore A hardness and surface roughness of two silicones for maxillofacial prosthetic treatment, under the influence of chemical disinfection and storage. Twenty-eight specimens were obtained, half of which were made of Silastic MDX 4-4210 silicone and, the other half were made of Silastic 732 RTV silicone. The specimens were divided into four groups...

متن کامل

Physical property comparison of 11 soft denture lining materials as a function of accelerated aging.

Soft denture-lining materials are an important treatment option for patients who have chronic soreness associated with dental prostheses. Three distinctly different types of materials are generally used. These are plasticized polymers or copolymers, silicones, or polyphosphazene fluoroelastomer. The acceptance of these materials by patients and dentists is variable. The objective of this study ...

متن کامل

A new method for photodynamic disinfection of prosthetic constructions and impressions in prosthetic dentistry.

INTRODUCTION Photodynamic therapy is a topical treatment of pathogens that involves the use of a photoactive dye (photosensitizer), which is non-toxic when not exposed to light and activated by light of a specific wavelength in the presence of oxygen. The highly cytotoxic oxygen species generated by the induced photophysical processes inactivate the pathogenic cells. The PURPOSE of this study w...

متن کامل

Effect of accelerated aging on the viscoelastic properties of a medical grade silicone.

The viscoelastic properties of cylinders (diameter 5 mm, height 2.2 ± 0.2 mm) of Nagor silicone elastomer of medium hardness, were investigated before and after the specimens had undergone accelerated aging in saline solution at 70°C for 38, 76 and 114 days (to simulate aging at 37°C, for 1, 2 and 3 years, respectively). All sets of specimens were immersed in physiological saline solution at 37...