The relevance of image quality indices for dose optimization in abdominal multi-detector row CT in children: experimental assessment with pediatric phantoms.

نویسندگان

  • H J Brisse
  • J Brenot
  • N Pierrat
  • G Gaboriaud
  • A Savignoni
  • Y De Rycke
  • S Neuenschwander
  • B Aubert
  • J-C Rosenwald
چکیده

This study assessed and compared various image quality indices in order to manage the dose of pediatric abdominal MDCT protocols and to provide guidance on dose reduction. PMMA phantoms representing average body diameters at birth, 1 year, 5 years, 10 years and 15 years of age were scanned in a four-channel MDCT with a standard pediatric abdominal CT protocol. Image noise (SD, standard deviation of CT number), noise derivative (ND, derivative of the function of noise with respect to dose) and contrast-to-noise ratio (CNR) were measured. The 'relative' low-contrast detectability (rLCD) was introduced as a new quantity to adjust LCD to the various phantom diameters on the basis of the LCD(1%) assessed in a Catphan phantom and a constant central absorbed dose. The required variations of CTDIvol(16) with respect to phantom size were analyzed in order to maintain each image quality index constant. The use of a fixed SD or CNR level leads to major dose ratios between extreme patient sizes (factor 22.7 to 44 for SD, 31.7 to 51.5 for CNR(2.8%)), whereas fixed ND and rLCD result in acceptable dose ratios ranging between factors of 2.9 and 3.9 between extreme phantom diameters. For a 5-9 mm rLCD1(%), adjusted ND values range between -0.84 and -0.11 HU mGy(-1). Our data provide guidance on dose reduction on the basis of patient dimensions and the required rLCD (e.g., to get a constant 7 mm rLCD(1%) for abdominal diameters of 10, 13, 16, 20 and 25 cm, tube current-time product should be adjusted in order to obtain CTDIvol(16) values of 6.2, 7.2, 8.8, 11.6 and 17.7 mGy, respectively).

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

An assessment of the Doses Received by Children from CT Examinations Along with the QC Parameters from a Conventional CT System

Introduction: In 2000, the UNSCEAR reported that CT constitutes 5% of all the medical x-ray  examinations and it contributes 34% of the resultant collective dose worldwide. Children are more  sensitive to the ionizing radiations than adults. So, routine quality control tests are expected to be  carried out periodically on the CT scanners. The aim of this research was to estimate the effective  ...

متن کامل

Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality

Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(me...

متن کامل

Optimization of Parameters in 16-slice CT-‌‌scan Protocols for Reduction of the Absorbed Dose

Introduction In computed tomography (CT) technology, an optimal radiation dose can be achieved via changing radiation parameters such as mA, pitch factor, rotation time and tube voltage (kVp) for diagnostic images. Materials and Methods In this study, the brain, abdomen, and thorax scaning was performed using Toshiba 16-slice scannerand standard AAPM and CTDI phantoms. AAPM phantom was used for...

متن کامل

Assessment and evaluation of patient doses in adult common CT examinations towards establishing national diagnostic reference levels

Background: Patient radiation doses from computed tomography (CT) are increasing due to the number of CT examinations performed every day. The aim of this study was assess and evaluate patient radiation doses for adult’s common CT examinations to derive local diagnostic guidance levels for common CT examinations. Materials and Methods: Volume and weighted computed tomography dose index (C...

متن کامل

Detection of low-contrast objects: experimental comparison of single- and multi-detector row CT with a phantom.

PURPOSE To compare single-detector row computed tomography (CT) and multi-detector row CT by using an experimental phantom containing a contrast-detail modulus for detection of low-contrast structures to optimize acquisition protocols. MATERIALS AND METHODS The parameters milliampere seconds, reconstructed section thickness, and elementary collimation for multi-detector CT were varied for two...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 54 7  شماره 

صفحات  -

تاریخ انتشار 2009