Comparison of ICare, dynamic contour tonometer, and ocular response analyzer with Goldmann applanation tonometer in glaucoma

نویسندگان

  • EVELIEN VANDEWALLE
  • SOFIE VANDENBROECK
  • INGEBORG STALMANS
  • THIERRY ZEYEN
چکیده

Purpose. To compare the intraocular pressure (IOP) readings taken by ICare, Pascal dynamic contour tonometer (DCT), and ocular response analyzer (ORA) with those taken by Goldmann applanation tonometer (GAT). We sought to evaluate the influence of central corneal thickness (CCT) on IOP measurements and to compare patients’ preferences for the four tonometers. Methods. In this prospective study, 93 eyes from 93 patients were examined. Patients were randomly divided into 4 groups to vary the order in which the tonometers were applied. CCT was measured with an ultrasound pachymeter. Results. The average CCT was 558±47.4 μm. The mean ± standard deviation IOP for GAT, ICare, DCT, and ORA (Goldmann-correlated IOP) (ORAg) were 15.1±4.8, 15.7±5.7, 18.3±5.1, and 18.3±6.6 mmHg, respectively. There was no significant difference between the mean IOP obtained with GAT and ICare (p=0.14). There was also no difference in IOP levels between the mean IOP obtained with DCT and ORA (p=0.26). There was no correlation between IOP measurements and CCT for the four instruments. Bland-Altman graphs showed disagreement between the measurements taken by GAT and the other instruments. There was no significant difference in patients’ preference among the 4 instruments (p=0.48). Conclusions. IOP readings from ICare were consistent with those from GAT, whereas DCT readings correspond well to ORAg measurements. DCT and ORA readings both overestimated the GAT readings. There was no correlation between the IOP measurements and the CCT for the 4 instruments. There was no significant difference in patients’ preference among the 4 instruments (Eur J Ophthalmol 2009 ; 19 : )

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of intraocular pressure measurement between rebound, non-contact and Goldmann applanation tonometry in treated glaucoma patients.

BACKGROUND To compare the intraocular pressure readings obtained with the iCare rebound tonometer and the 7CR non-contact tonometer with those measured by Goldmann applanation tonometry in treated glaucoma patients. DESIGN A prospective, cross-sectional study was conducted in a private tertiary glaucoma clinic. PARTICIPANTS OR SAMPLES One hundred nine (54 males : 55 females) patients includ...

متن کامل

Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

PURPOSE (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer. METHODS A...

متن کامل

Erratum: A cross-sectional study to compare intraocular pressure measurement by sequential use of Goldman applanation tonometry, dynamic contour tonometry, ocular response analyzer, and Corvis ST

OBJECTIVE To study the correlation and effect of sequential measurement of intraocular pressure (IOP) with Goldmann applanation tonometer (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST. SETTING AND DESIGN Observational cross-sectional series from the comprehensive clinic of a tertiary eye care center seen during December 2012. METHODS One hundred and t...

متن کامل

Diagnostic accuracy of the iCare rebound tonometer compared to the Perkins applanation tonometer in assessing intraocular pressure in rural patients.

BACKGROUND Vision health is recognized as a critical unmet need in North America. The ocular morbidity associated with glaucoma results from increased intraocular pressure (IOP) and early detection is crucial for the management of glaucoma. Our objective was to find a diagnostically accurate screening tool for intraocular hypertension that can be used in rural communities. We sought to validate...

متن کامل

A comparison between rebound and Goldmann's tonometers in screening of patients for glaucoma.

Goldmann applanation tonometer (GAT) has been the gold standard for intraocular pressure (IOP) measurement for over five decades. However, the last few years have seen lots of research in an effort to develop the ultimate tonometer. One of the outcomes of that effort is the rebound tonometer (RBT) (Icare; Tiolat Oy, Helsinki, Finland), which has appeared in clinical practice after extensive res...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2009