A portable device based on the interrupter technique to measure bronchodilator response in schoolchildren.

نویسندگان

  • P D Bridge
  • H Lee
  • M Silverman
چکیده

A new device for measuring airway resistance following brief airflow interruption (Microlab 4000; Micromedical Ltd, UK) was evaluated in 25 asthmatic school children in comparison with well-established methods. Airway resistance was measured during brief airflow interruption (Rint), before and after administration of salbutamol 200 micrograms by metered-dose inhaler, and in the spirometric parameters, forced expiratory volume in one second (FEV1) and peak expiratory flow (PEF), and total respiratory system resistance at 6 Hz (Rrs,6) measured by the forced oscillation technique (FOT). The sensitivity index (SI) (mean change/baseline standard deviation) was calculated for each subject. At baseline, interrupter conductance, the reciprocal of Rint, correlated well with FEV1 (r = 0.837; p < 0.001) and PEF (r = 0.773; p < 0.001), and Rint correlated highly with Rrs,6 (r = 0.942; p < 0.001). The median intrasubject coefficient of variation of the interrupter method was higher than the FOT or either spirometric parameter; Rint 11%, Rrs,6 9%, FEV1 5% and PEF 5%. However, the sensitivity to detect change after bronchodilator, expressed as the median SI, did not differ significantly between measurements: Rint 3.5, Rrs,6 3.6, FEV1 2.4 and PEF 3.0. A significant response (SI > 2) was shown by the interrupter in 22 of the subjects compared with 16 by FEV1. The interrupter technique is useful for assessing changes in airway calibre in asthmatic school children, with a sensitivity at least as good as standard methods. Such a device could be of particular value in those too young to perform spirometry.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Using the interrupter technique to evaluate airway resistance in cystic fibrosis patients.

OBJECTIVE To measure airway resistance with the interrupter resistance (Rint) technique in patients with cystic fibrosis (CF) and to determine whether Rint values correlate with spirometric parameters, as well as to evaluate the accuracy of the Rint technique in determining the airway response to a bronchodilator. METHODS This was a cross-sectional study involving 38 children and adolescents ...

متن کامل

Evaluation of a new interrupter device for measuring bronchial responsiveness and the response to bronchodilator in 3 year old children.

The interrupter technique for measuring airway resistance is non-invasive and convenient, and therefore ideally suited for the assessment of induced changes in airway calibre in preschool children. The aim of this study was to evaluate a commercially available interrupter device (based on Microlab 4000), which calculates the interrupter resistance (Rint) from pressure and flow following a brief...

متن کامل

A portable device based on the interrupter technique for measuring airway resistance in preschool children.

The interrupter technique for measuring airway resistance is a noninvasive method reported to require minimal subject cooperation. Therefore it has a good potential for use in young children who are not able to cooperate with conventional lung function tests. The interrupter method is based on transient interruption of airflow at the mouth for a brief period during which alveolar pressure equil...

متن کامل

Design and Fabrication of a Portable 1-DOF Robotic Device for Indentation Tests

There are many tactile devices for indentation examinations to measure mechanical properties of tissue. The purpose of this paper is to develop a portable indentation robotic device to show its usability for measuring the mechanical properties of a healthy abdominal tissue. These measurements will help to develop suitable mathematical models representing abdominal tissue. A 1-DOF portable robot...

متن کامل

Measurement of airway resistance using the interrupter technique in preschool children in the ambulatory setting.

This study describes the feasibility, repeatability, and interrater reliability of the measurement of airway resistance by the interrupter technique (Rint) in children 2-5 yrs of age, and examines whether reversibility to bronchodilator can be demonstrated in wheezy children. The mean of six Rint values was taken as a measurement. If subjects could complete one measurement and then a second 15 ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The European respiratory journal

دوره 9 7  شماره 

صفحات  -

تاریخ انتشار 1996