Accuracy, precision and linearity of the portable flow-volume meter Microspiro HI-298.

نویسندگان

  • E Dompeling
  • C P van Schayck
  • H Folgering
  • H J van den Hoogen
  • C van Weel
چکیده

The accuracy, precision and linearity of a new portable flow-volume meter, the Microspiro HI-298 (Chest Corporation, Tokyo, Japan), was investigated using a Fleisch no. 4 pneumotachograph as a standard. After connection and calibration of the pneumotachograph and the Microspiro, a healthy subject performed 44 forced vital capacity (FVC) manoeuvres at different levels of lung inflation. The FVC of these expirations ranged from 2.5-5.1 l. Linear regression of Microspiro values (dependent variable) on Fleisch pneumotachograph values (independent variable) showed that a good linear relationship existed: Pearson correlation coefficients ranged from 0.938-0.985. Linearity of the Microspiro was good except for the peak expiratory flow rate (PEFR) and the maximal expiratory flow at 25% of the expired volume (MEF75). The random error (measure of precision) of all flow-volume (F-V) indices was lower than 5%. The systematic error (measure of accuracy) was low for the forced expiratory volume in one second (FEV1) and the FVC (1% and 4.6%, respectively) but much higher for the instantaneous expiratory flows (PEFR 11.0%; MEF75 7.0%; MEF50 8.5%; MEF25 11.4%). Only the total error in FEV1 complied with the tolerance of 4% of the European Community for Coal and Steel (ECCS). When the measured values were adjusted according to the regression equations of this study, all F-V indices were accurate and precise within 5%. It was concluded that the portable Microspiro HI-298 is a useful instrument for bedside, work-site spirometry and for use in general practice.(ABSTRACT TRUNCATED AT 250 WORDS)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

HIGH-PERFORMANCE LIQUID CHROMATOGRAPHIC DETERMINATION OF THEOPHYLLINE IN HUMAN SERUM

A simple, rapid and sensitive HPLC method for the determination of theophylline (T) in human serum has been developed. An isocratic system consisting of a /-l Bondapak C18 column, mobile phase of methanol, phosphate buffer (22:78, pH=4.5), and a flow rate of 1.4 mL/min was used. The eluent was detected by UV at 275 nm at room temperature. 8-Chlorotheophylline (8-CT) was used as an internal ...

متن کامل

Assessment of accuracy and applicability of a new electronic peak flow meter and asthma monitor.

The aim of this study was to assess the accuracy and applicability of a portable electronic peak flow meter combined with an asthma monitor (AM1, Jaeger, Germany) which measures peak expiratory flow (PEF), forced expiratory volume in one second (FEV1) and forced vital capacity (FVC). The technical accuracy in PEF, FEV1 and FVC measurement was tested according to American Thoracic Society (ATS) ...

متن کامل

Assessment of accuracy and applicability of a portable electronic diary card spirometer for asthma treatment.

A pocked-sized turbine flowmeter and spirometer device, integrated with an electronic diary card (EDC-spirometer, Micro Medical, U.K.), was tested with a mechanical calibrator, in an outpatient clinic and in the home situation. A screen pneumotachometer was used as flow and volume reference. Ten devices were tested; interdevice variability was small with a mean variation coefficient of 1.1% for...

متن کامل

Rapid Analysis of Glibenclamide Using an Environmentally Benign Stability-Indicating RP-HPLC Method

An environmentally benign RP-HPLC approach for rapid analysis of glibenclamide in pure form, developed nanoemulsion and commercial tablets was developed and validated in present investigation. The green chromatographic identification was performed on Lichrosphere 250 X 4.0 mm RP C8 column having a 5µm packing as a stationary phase using a combination of ethanol: methanol (50:50 % v/v) as a mobi...

متن کامل

Rapid Analysis of Glibenclamide Using an Environmentally Benign Stability-Indicating RP-HPLC Method

An environmentally benign RP-HPLC approach for rapid analysis of glibenclamide in pure form, developed nanoemulsion and commercial tablets was developed and validated in present investigation. The green chromatographic identification was performed on Lichrosphere 250 X 4.0 mm RP C8 column having a 5µm packing as a stationary phase using a combination of ethanol: methanol (50:50 % v/v) as a mobi...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 4 5  شماره 

صفحات  -

تاریخ انتشار 1991