Characterization of a fiber-coupled Al2O3:C luminescence dosimetry system for online in vivo dose verification during Ir brachytherapy
نویسندگان
چکیده
A prototype of a new dose-verification system has been developed to facilitate prevention and identification of dose delivery errors in remotely afterloaded brachytherapy. The system allows for automatic online in vivo dosimetry directly in the tumor region using small passive detector probes that fit into applicators such as standard needles or catheters. The system measures the absorbed dose rate 0.1 s time resolution and total absorbed dose on the basis of radioluminescence RL and optically stimulated luminescence OSL from aluminum oxide crystals attached to optical fiber cables 1 mm outer diameter . The system was tested in the range from 0 to 4 Gy using a solidwater phantom, a Varian GammaMed Plus Ir PDR afterloader, and dosimetry probes inserted into stainless-steel brachytherapy needles. The calibrated system was found to be linear in the tested dose range. The reproducibility one standard deviation for RL and OSL measurements was 1.3%. The measured depth-dose profiles agreed well with the theoretical expectations computed with the EGSNRC Monte Carlo code, suggesting that the energy dependence for the dosimeter probes relative to water is less than 6% for source-to-probe distances in the range of 2–50 mm. Under certain conditions, the RL signal could be greatly disturbed by the so-called stem signal i.e., unwanted light generated in the fiber cable upon irradiation . The OSL signal is not subject to this source of error. The tested system appears to be adequate for in vivo brachytherapy dosimetry. © 2009 American Association of Physicists in Medicine. DOI: 10.1118/1.3063006
منابع مشابه
Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy.
PURPOSE The purpose of this study is to present and evaluate a dose-verification protocol for pulsed dose-rate (PDR) brachytherapy based on in vivo time-resolved (1 s time resolution) fiber-coupled luminescence dosimetry. METHODS Five cervix cancer patients undergoing PDR brachytherapy (Varian GammaMed Plus with 192Ir) were monitored. The treatments comprised from 10 to 50 pulses (1 pulse/h) ...
متن کاملDeveloping a Verification and Training Phantom for Gynecological Brachytherapy System
Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...
متن کاملRectal and Bladder Dose Measurements in the Intracavitary Applications of Cervical Cancer Treatment with HDR Afterloading System: Comparison of TPS Data with MOSFET Detector
Background: Intracavitary brachytherapy plays a major role in management of cervical carcinoma. Assessment of dose received by OAR’s therefore becomes crucial for the estimation of radiation toxicities in HDR brachytherapy. Objective: Purpose of this study is to evaluate the role of in vivo dosimetry in HDR brachytherapy and to compare actual doses delivered to OAR’ s with those calculate...
متن کاملA Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy
Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...
متن کاملAn Investigation of the Dose Distribution from LDR Ir-192 Wires in the Triangular Implants of the Paris System using Polymer Gel Dosimetry
Introduction: Polymer gels are modern dosimeters providing three dimensional dose distributions. These dosimeters can be used in brachytherapy in which the tumor dimension is relatively small and the dose gradient is high. In this study, the ability of the MAGICA polymer gel was investigated for assessing the absolute dose values as well as the dose distribution of low dose rate (LDR) Ir-192 wi...