Harris-Galante II acetabular cup: a survival analysis.
نویسندگان
چکیده
PURPOSE To determine the survival rate of the Harris-Galante (HG) II cup and identify factors contributing to failure. METHODS 271 primary total hip arthroplasties (THAs) performed between 1992 and 1999 in our hospitals using the HG II acetabular cup were retrospectively reviewed due to a high incidence of liner dissociation. RESULTS 48 THAs were revised, 32 (67%) of which were due to liner dissociation. The Kaplan-Meier 10-year survival rate was 73%. Dislocation (p<0.001) and acetabular size (p=0.042) were independent predictors of acetabular component failure in the Cox's regression model. CONCLUSION The long-term outcome of the HG II cup is unsatisfactory. Although the HG II cup is no longer produced, numerous patients have this prosthesis in situ and should be reviewed regularly, as liner dissociation and retroacetabular osteolysis may occur.
منابع مشابه
Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups.
Locking mechanisms and metal-liner interface surfaces of six modular acetabular systems were evaluated to determine their effect on micromotion and backside wear of the polyethylene liner. Rotational and axial motion between the metal shell and polyethylene liner was measured in the Duraloc (DePuy, Warsaw, IN), Harris-Galante (Zimmer, Warsaw, IN), Impact (Biomet, Warsaw, IN), Lip Loc (Biomet), ...
متن کاملSurvival analysis of the Harris-Galante I acetabular cup.
We reviewed 264 consecutive primary total hip replacements in 244 patients in which the Harris-Galante I porous-coated acetabular component had been used. The mean follow-up was 95 months (69 to 132). In 221 arthroplasties a 32 mm ceramic head had been used, and in the other 43 one of cobalt-chrome alloy. There were 124 women and 120 men with a mean age at operation of 56.8 years (21 to 83). Su...
متن کاملCementless Hip Arthroplasty in Southern Iran, Midterm Outcome and Comparison of Two Designs
Background: Cementless hip prosthesis was designed to provide biologic fixation, without the use of cement. The second generation components have shown more reliable bone ingrowths and survival rates. We are reporting a midterm result of two designs of cementless prosthesis in a unique culture with different social habits and expectations.Methods: 52 primary cementless total hip arthroplasty in...
متن کاملHarris Galante cementless acetabular replacement in avascular necrosis.
We describe the results of 76 total arthroplasties of the hip for stage-III or stage-IV avascular necrosis of the femoral head. Harris Galante Porous cups were used in 63 patients between 1986 and 1994 and followed prospectively. We reviewed 70 hips with a follow-up of more than five years (mean 7.6). At the latest review the mean Harris Hip Score had improved from a preoperative value of 29 +/...
متن کاملIncreased risk of revision of acetabular cups coated with hydroxyapatite
BACKGROUND Hydroxyapatite (HA) is the main inorganic component of bone, and HA coating is widely used on acetabular cups in hip arthroplasty. It has been suggested that this surface finish improves cup survival. METHODS All patients registered in the Swedish Hip Arthroplasty Register between 1992 and 2007 with an uncemented acetabular implant that was available either with or without HA coati...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of orthopaedic surgery
دوره 16 2 شماره
صفحات -
تاریخ انتشار 2008