Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization.

نویسندگان

  • Cheuk-Man Yu
  • Li Wang
  • Elaine Chau
  • Raymond Hon-Wah Chan
  • Shun-Ling Kong
  • Man-Oi Tang
  • Jill Christensen
  • Robert W Stadler
  • Chu-Pak Lau
چکیده

BACKGROUND Patients with heart failure are frequently hospitalized for fluid overload. A reliable method for chronic monitoring of fluid status is therefore desirable. We evaluated an implantable system capable of measuring intrathoracic impedance to identify potential fluid overload before heart failure hospitalization and to determine the correlation between intrathoracic impedance and standard measures of fluid status during hospitalization. METHODS AND RESULTS Thirty-three patients with NYHA class III and IV heart failure were implanted with a special pacemaker in the left pectoral region and a defibrillation lead in the right ventricle. Intrathoracic impedance was regularly measured and recorded between the lead and the pacemaker case. During hospitalizations, pulmonary capillary wedge pressure and fluid status were monitored. Ten patients were hospitalized for fluid overload 25 times over 20.7+/-8.4 months. Intrathoracic impedance decreased before each admission by an average of 12.3+/-5.3% (P<0.001) over an average of 18.3+/-10.1 days. Impedance reduction began 15.3+/-10.6 days (P<0.001) before the onset of worsening symptoms. There was an inverse correlation between intrathoracic impedance and pulmonary capillary wedge pressure (r=-0.61, P<0.001) and between intrathoracic impedance and net fluid loss (r=-0.70, P<0.001) during hospitalization. Automated detection of impedance decreases was 76.9% sensitive in detecting hospitalization for fluid overload, with 1.5 false-positive (threshold crossing without hospitalization) detections per patient-year of follow-up. CONCLUSIONS Intrathoracic impedance is inversely correlated with pulmonary capillary wedge pressure and fluid balance and decreased before the onset of patient symptoms and before hospital admission for fluid overload. Regular monitoring of impedance may provide early warning of impending decompensation and diagnostic information for titration of medication.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Continuous monitoring of intrathoracic impedance and right ventricular pressures in patients with heart failure.

BACKGROUND Hemodynamic monitoring using implantable devices may provide early warning of volume overload in patients with heart failure (HF). This study was designed to prospectively compare information from intrathoracic impedance monitoring and continuous right ventricular pressure measurements in patients with HF. METHODS AND RESULTS Sixteen patients with HF (age, 63.5+/-13.8 years; left v...

متن کامل

Decreased Intrathoracic Impedance Associated With OptiVol Alert Can Diagnose Increased B-Type Natriuretic Peptide

indicators even with day-to-day follow-up of patients with HF.3,4 Daily monitoring of the thoracic fluid status with intrathoracic impedance (ITI) is a diagnostic feature of implantable cardioverter defibrillators (ICD) or can be carried out in combination with cardiac resynchronization therapy (CRT-D). The monitoring of ITI may provide helpful information for the early detection of HF decompen...

متن کامل

Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure.

AIMS To evaluate the utility of intrathoracic impedance monitoring for detecting heart failure (HF) deterioration in patients with an implanted cardiac resynchronization/defibrillation device. METHODS AND RESULTS Patients enrolled in the European InSync Sentry Observational Study were audibly alerted by a device algorithm if a decrease in intrathoracic impedance suggested fluid accumulation. ...

متن کامل

Heart Failure Intrathoracic Impedance Monitoring, Audible Patient Alerts, and Outcome in Patients With Heart Failure

Background—Heart failure is associated with frequent hospitalizations, often resulting from volume overload. Measurement of intrathoracic impedance with an implanted device with an audible patient alert may detect increases in pulmonary fluid retention early. We hypothesized that early intervention could prevent hospitalizations and affect outcome. Methods and Results—We studied 335 patients wi...

متن کامل

Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure.

BACKGROUND Heart failure is associated with frequent hospitalizations, often resulting from volume overload. Measurement of intrathoracic impedance with an implanted device with an audible patient alert may detect increases in pulmonary fluid retention early. We hypothesized that early intervention could prevent hospitalizations and affect outcome. METHODS AND RESULTS We studied 335 patients ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Circulation

دوره 112 6  شماره 

صفحات  -

تاریخ انتشار 2005