Evaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System

نویسندگان

  • Mamta Mahur
  • Om Prakash Gurjar
  • RK Grover
  • PS Negi
  • Richa Sharma
  • Anshu Singh
  • Munendra Singh
چکیده

Article type: Original Article Introduction: In radiotherapy treatment planning system (TPS), basic input is the data from computed tomography (CT) scan, which takes into account the effect of inhomogeneities in dose calculations. Measurement of CT numbers may be affected by scanner-specific parameters. Therefore, it is important to verify the effect of different CT scanning protocols on Hounsfield unit (HU) and its impact on dose calculation. This study was carried out to analyse the effect of different tube voltages on HU for various tissue substitutes in phantom and their dosimetric impact on dose calculation in TPS due to variation in HU–relative electron density (RED) calibration curves. Materials and Methods: HU for different density materials was obtained from CT images of the phantom acquired at various tube voltages. HU-RED calibration curves were drawn from CT images with various tissue substitutes acquired at different tube voltages used to quantify the error in dose calculation for different algorithms. Doses were calculated on CT images acquired at 120 kVp and by applying CT number to RED curve obtained from 80, 100, 120, and 140 kVp voltages. Results: No significant variation was observed in HU of different density materials for various kVp values. Doses calculated with applying different HU-RED calibration curves were well within 1%. Conclusion: Variation in doses calculated by algorithms with various HU-RED calibration curves was found to be well within 1%. Therefore, it can be concluded that clinical practice of using the standard HU-RED calibration curve by a 120 kVp CT acquisition technique is viable. Article history: Received: Feb 11, 2017 Accepted: May 01, 2017

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Effect of Different Computed Tomography Scanning Protocols on Hounsfield Unit and Its Impact on Dose Calculation by Treatment Planning System

Introduction: In radiotherapy treatment planning system (TPS), basic input is the data from computed tomography (CT) scan, which takes into account the effect of inhomogeneities in dose calculations. Measurement of CT numbers may be affected by scanner-specific parameters. Therefore, it is important to verify the effect of different CT scanning protocols on Hounsfield unit (HU) and its impact o...

متن کامل

Evaluation of the gray level in CBCT systems and its relationship with HU in CT Scanners

Introduction: Cone-beam CT (CBCT) is an imaging system which offers three-dimensional (3D), multiplanar images and has many advantages over computed tomography (CT) including shorter acquisition times for the resolution desired in dentistry, lower radiation dose to the patient, reasonable price and higher spatial resolution but CBCT scanners are unable to display actual Hounsf...

متن کامل

Investigating the effects of different kernels used for CT image reconstruction on dose distributions in treatment planning of kidney cancer radiotherapy

Introduction: The quality of CT images used for treatment planning of cancer patients is an important issue in accurate outlining of the tumor volume and organs at risk. Different kernels in CT scanner systems are available for improving the image quality. Applying these kernels  on   CT  images  will  change   the   CT  numbers  and  electron   density   of   tissues, conse...

متن کامل

Comparisons of Hounsfield Unit Linearity between Images Reconstructed using an Adaptive Iterative Dose Reduction (AIDR) and a Filter Back-Projection (FBP) Techniques

Background: The HU linearity is an essential parameter in a quantitative imaging and the treatment planning systems of radiotherapy. Objective: This study aims to evaluate the linearity of Hounsfield unit (HU) in applying the adaptive iterative dose reduction (AIDR) on CT scanner and its comparison to the filtered back-projection (FBP).Material and Methods: In this experimental phan...

متن کامل

Dosimetric accuracy of the cone‐beam CT‐based treatment planning of the Vero system: a phantom study

We report an investigation on the accuracy of dose calculation based on the cone-beam computed tomography (CBCT) images of the nonbowtie filter kV imaging system of the Vero linear accelerator. Different sets of materials and tube voltages were employed to generate the Hounsfield unit lookup tables (HLUTs) for both CBCT and fan-beam CT (FBCT) systems. The HLUTs were then implemented for the dos...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2017