Pacemaker Failure Due to Loss of Fluid Seal in a Patient with a 5/6mm Pacemaker Header Port and a 5mm Unipolar Lead

نویسندگان

  • Paul Chun Yih Lim
  • Boon Yew Tan
  • Kah Leng Ho
  • Chee Keong Ching
  • Wee Siong Teo
چکیده

INTRODUCTION Before IS-1 (3.2 mm) standardization of pacemaker leads and connectors, 5/6 mm connector ports accomodated 5mm or 6mm diameter lead connector pins. CASE REPORT A patient with sick sinus syndrome underwent implantation of a 5mm unipolar atrial lead, mated to a 5/6mm connector port Medtronic Spectrax Sx 5985 pacemaker. Pulse generator reached ERI in 2006, with change out to a Medtronic Sigma SSR306 (5/6mm connector port) and preservation of the 5mm lead. She was admitted in 2010 for atrial lead non capture from blood leak and corrosion of the header-connector pin apparatus. DISCUSSION 5/6mm pacemaker header ports have a 5mm flexible sealing ring at the port entrance to seal 5mm or 6mm lead connector pins. The inner barrel diameter of the connector port is 6mm and insertion of a 5mm lead results in a 0.5mm tolerance circumferentially. Should the seal be compromised, blood can corrode the apparatus. To minimize this, we can employ (a) a cinching tie to further seal the silicone ring (b) universal adaptor sleeves (c) splice kits (d) lead adaptor kits. Aging leads, adaptor kits or sleeves themselves can result in lead failure. It may be safer to re-implant the entire system. CONCLUSIONS A 5/6mm configuration pacemaker header connector port allows for significant tolerances when a 5 mm lead is used. Consideration must be made to prevent leaks.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pacemaker Twiddler's Syndrome

A 76-yr-old man with unipolar ventricular pacemaker, previously implanted in the right ventricular apex, presented with syncopal attack and twitching of the right pectoralis muscle. He finally was diagnosed as having Twiddler's syndrome. The patient had a pulse rate of 36/min and cannon waves in the neck veins. Chest radiography showed a dislodged pacing lead completely wrapped around the pulse...

متن کامل

Pacemaker lead endocarditis with hiccups (Kalayci)

Background: Lead-related infections that might develop after pacemaker implantation associated with high mortality and morbidity rates are challenging to manage and pose high-cost. Patients with lead-related infections usually present with fever, chills and fatigue and the treatment can be challenging unless the implant system is extracted. Case presentation: A 66-year old male patient who und...

متن کامل

The Improvement of the Atrial Flutter Rhythm upon the Removal of the Infected Permanent Pacemaker Lead

Pacemaker infection has multiple risk factors. Its presentation is most often similar to infected endocarditis and the diagnosis is made through studying blood cultures. Transesophageal echocardiography can confirm the diagnosis. The most common microorganisms are staphylococcus speciesis. As a matter of fact, complete pacemaker removal appears to be the only definite treatment. We presented a ...

متن کامل

The influence of neutron contamination on pacemaker in photon beam radiotherapy by LINAC using the Monte Carlo method

Introduction: In radiation therapy with high-energy photon beams (E > 7 MeV) neutrons are generated mainly in LINACs head thorough (γ, n) interactions. These neutrons affect the shielding requirements in radiation therapy rooms. According to the protocol TG-34, photon absorbed dose of 10Gy can cause permanent damage to the pacemaker and the dose of 2Gy can make minor changes in...

متن کامل

بررسی مشکلات جسمی، روانی و اجتماعی بیماران دارای دستگاه ضربا ساز قلب

  The aim of this cross sectional study is the assessment of physical and psychosocial problems of patients with pacemaker and effects of social - individual factors (such as age, sex, marital status, level of education,....) on the intensity of problems. In the study 60 patients (27 women, 33 men) that their mean age was 60.07 years, were studied. Sampling was purposive and the samples were se...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 13  شماره 

صفحات  -

تاریخ انتشار 2013