Temperature dependence of led and its theoretical effect on pulse oximetry.

نویسندگان

  • K J Reynolds
  • J P de Kock
  • L Tarassenko
  • J T Moyle
چکیده

Ambient temperature is known to affect the emission spectrum of a light-emitting diode (LED). This study has investigated the effect of changes in ambient temperature on the emission spectra of two LED with peak emission wavelengths similar to those used in pulse oximetry. There was a 5.5-nm increase in the peak wavelength for a 660-nm LED, and a 7.8-nm increase in the peak wavelength for a 950-nm LED as temperature increased from 0 to 50 degrees C. Using a simple theoretical model based on the Beer-Lambert law, the effect of these shifts in wavelength on pulse oximeter accuracy was examined and found to be negligible over the temperature range studied.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation and Improvement of Knowledge of Medical staff on Pulse Oximetry in Educational Hospitals

Background: Pulse Oximeter monitoring has become so common over previous decades that in some health care units, blood oxygen level is now considered as the fifth vital sign. However, it seems there is still some lack of knowledge and/or experience among medical staff operating pulse oximeter and interpreting its results. The aim of this study was to investigate and compare the knowledge on Pul...

متن کامل

Evaluation and Improvement of Knowledge of Medical staff on Pulse Oximetry in Educational Hospitals

Background: Pulse Oximeter monitoring has become so common over previous decades that in some health care units, blood oxygen level is now considered as the fifth vital sign. However, it seems there is still some lack of knowledge and/or experience among medical staff operating pulse oximeter and interpreting its results. The aim of this study was to investigate and compare the knowledge on Pul...

متن کامل

Effect of Nail Polish on Pulse Oximetry Findings in Healthy Volunteers: A Randomized Clinical Trail

Background and Objectives: Pulse oximetry is the most common technique for monitoring hemoglobin oxygen saturation (SpO2). Different colors and brands of nail polish may cause disturbance in the reading and interpretation of oxygen saturation. The aim of this study was to determine the effect of different colors of nail polish on oxygen saturation measured by pulse oximeter. Methods: Thirty he...

متن کامل

Evaluation of Pulse Oximetry in the Early Detection of Congenital Heart Diseases in Newborns

Background: Congenital heart diseases, which are asymptomatic at birth, are the most important causes of infant mortality. This study aimed to evaluate the role of pulse oximetry in the early detection of congenital heart diseases among newborns.Methods: In this cross-sectional descriptive study, 1230 newborns who were born in university hospitals in an urban area of Iran were placed unde...

متن کامل

A Three-Wavelength Pulse Oximeter for Carboxyhemoglobin Determination

Pulse oximetry measures the relative concentration of arterial oxygenated hemoglobin and reduced hemoglobin. This paper presents the design of a pulse oximeter, which measures the presence of a third substance, carboxyhemoglobin (HbCO), by unfiltered LEDs at 660 nm, 810 nm, and 940 nm. Background information is given on the theory of pulse oximetry. Theoretical approximations and calculations w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British journal of anaesthesia

دوره 67 5  شماره 

صفحات  -

تاریخ انتشار 1991