In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening

نویسندگان

  • Onur Dincer Kose
  • Burcin Karataslı
  • Sabit Demircan
  • Taha Emre Kose
  • Erhan Cene
  • Serhan Aydın Aya
  • Mehmet Ali Erdem
  • Abdulkadir Burak Cankaya
چکیده

Preload is applied to screws manually or using a torque wrench in dental implant systems, and the preload applied must be appropriate for the purpose. The aim of this study was to assess screw loosening and bending/torsional moments applied by clinicians of various specialties following application of manual tightening torque to combinations of implants and abutments. Ten-millimeter implants of 3.7 and 4.1 mm diameters and standard or solid abutments were used. Each group contained five implant-abutment combinations. The control and experimental groups comprised 20 and 160 specimens, respectively. Implants in the experimental group were tightened by dentists of different specialties. Torsional and bending moments during tightening were measured using a strain gauge. Control group and implants with preload values close to the ideal preload were subjected to a dynamic loading test at 150 N, 15 Hz, and 85,000 cycles. The implants that deformed in this test were examined using an optical microscope to assess deformities. Manual tightening did not yield the manufacturer-recommended preload values. Dynamic loading testing suggested early screw loosening/fracture in samples with insufficient preload.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Manual maximum closing torque in three implant abutment systems

Background and Aims: Diferences whithin shape and size of hand wrenches in diferent systems could affect the torque applied. Therefore, evaluation of wrenches has been emphasized. The aim of the present study was to compare the torque values of the long and short torque wrenches in different implant systems including ITI, Implantium and Nobel Biocare. Materials and Methods: In this experimenta...

متن کامل

Comparison of the Effect of Two Different Abutment Diameters on Detorque Value Using Cyclic Loading and Thermocycling

Background and Aim: Screw loosening is a common problem with both screw-retained and cemented implant restorations. It is assumed that the abutment diameter affects detorque value and screw loosening. We aimed to determine the effect of two different abutment diameters on detorque value using cyclic loading and thermocycling. Materials and Methods: This in-vitro experimental study was conducte...

متن کامل

Effect of Abutment Connection Type and Cyclic Loading on Removal Torque Value

Background and Aim: Maintaining the primary torque of the abutment screw is a common problem related to implant-supported restorations; a failure in this respect often leads to screw loosening. The abutment screw plays a critical role in the long-term stability of restorations. This study aimed to assess the effect of two abutment connection types with and without cyclic loading on removal torq...

متن کامل

Torque Removal Evaluation of Screw in One-Piece and Two-Piece Abutments Tightened with a Handheld screwdriver

Introduction: Some clinicians use a handheld screw driver instead of a torque wrench to definitively tighten abutment screws. The aim of this study was to compare the removal torque of one-piece and two-piece abutments tightened with a handheld driver and a torque control ratchet. Methods: 40 ITI implants were placed in acrylic blocks and divided into 4 groups. In groups one and two, 10 ITI one...

متن کامل

Preload and Torque Removal Evaluation of Three Different Abutment Screws for Single Standing Implant Restorations

Several authors still consider the mechanical problems of fracture and component loosening as the main causes of failure of implant-supported restorations. The purpose of this in vitro study was to compare the preload of three types of screw for transmucosal abutment attachment used in single implant-supported prosthesis through strain gauge and removal torque measurements. Three external hex f...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2017  شماره 

صفحات  -

تاریخ انتشار 2017