Cone beam computed tomography number errors and consequences for radiotherapy planning: an investigation of correction methods.
نویسندگان
چکیده
PURPOSE The potential of keV cone beam computed tomography (CBCT) for guiding adaptive replanning is well-known. There are impediments to this, one being CBCT number accuracy. The purpose of this study was to investigate CBCT number correction methods and the affect of residual inaccuracies on dose deposition. Four different correction strategies were applied to the same patient data to compare performance and the sophistication of correction-method needed for acceptable dose errors. METHODS AND MATERIALS Planning CT and CBCT reconstructions were used for 12 patients (6 brain, 3 prostate, and 3 bladder cancer patients). All patients were treated using Elekta linear accelerators and XVI imaging systems. Two of the CBCT number correction methods investigated were based on an algorithm previously proposed by the authors but only previously applied to phantoms. Two further methods, based on an approach previously suggested in the research literature, were also examined. Dose calculations were performed using scans of a "worst" subset of patients using the Pinnacle³ version 9.0 treatment planning system and the patients' clinical plans. RESULTS All mean errors in CBCT number were <50 HU, and all correction methods performed well or adequately in dose calculations. The worst single dose discrepancy identified for any of the examined methods or patients was 3.0%. Mean errors in the doses to treatment volumes or organs at risk were negatively correlated with the mean error in CT number. That is, a mean CT number that was too large, averaged over the entire CBCT volume, implied an underdosing in a volume-of-interest and vice versa. CONCLUSIONS Results suggest that (1) the correction of CBCT numbers to within a mean error of 50 HU in the scan volume provides acceptable discrepancies in dose (<3%) and (2) this is achievable with even quite unsophisticated correction methods.
منابع مشابه
Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality
Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...
متن کاملDosimetric Effects of Rotational Setup Error in Volumetric Modulated Arc Radiotherapy on Brain Tumor Patients
Introduction: This study examined the dosimetric effects based on the rotational setup error to correct patient setup errors occur during volumetric modulated arc radiotherapy (VMAT) for brain tumor patients. Material and Methods: This study included 1129 cases of cone beam computed tomography (CBCT) images obtained from 46 brain tumor patients, who experienced VMAT and used the 6DoF (degree o...
متن کاملUsability assessment of cone beam computed tomography with a full-fan mode bowtie filter compared to that with a half-fan mode bowtie filter
Background: In intensity modulated radiation therapy, cone beam computed tomography (CT) has been used to evaluate patients prior to treatment. This study conducted a comparative evaluation of the image reconstruction ability of the clinically used half-fan bowtie filter and the full-fan bowtie filter. Materals and Methods: A CT simulation marker was inserted inside a human phantom, and the pel...
متن کاملMethods to evaluate the performance of kilovoltage cone-beam computed tomography in the three-dimensional reconstruction space
Background: Cone-beam computed tomography (CBCT) scanners for image-guided radiotherapy are in clinical use today, but there has been no consensus on uniform acceptance to verify the CBCT image quality yet. The present work proposed new methods to fully evaluate the performance of CBCT in its three-dimensional (3D) reconstruction space. Materials and Methods: Compared to the traditional methods...
متن کاملCorrection to: An Ex vivo Analysis of Apical Transportation in Root Canals Prepared with BT Race Rotary and Hand K-Flexofile by Cone Beam Computed Tomography
Correction to: An Ex vivo Analysis of Apical Transportation in Root Canals Prepared with BT Race Rotary and Hand K-Flexofile by Cone Beam Computed Tomography Abbas Mesgarani1, Azadeh Zakariaie2, Azam Haddadi Kohsar1, Sina Haghanifar3, Jamshid Yazdani Charati4, Arman Ebrahimi5 1 Assistant Professor, Department of Endodontics, Faculty of Dentistry, Mazandaran University of Medical S...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- International journal of radiation oncology, biology, physics
دوره 84 1 شماره
صفحات -
تاریخ انتشار 2012