A digital filter derived from linear discriminant analysis (LDA) is developed for recovering impulse responses in photon counting from a high speed photodetector (rise time of ~1 ns) and applied to remove ringing distortions from impedance mismatch in multiphoton fluorescence microscopy. Training of the digital filter was achieved by defining temporally coincident and noncoincident transients and identifying the projection within filter-space that best separated the two classes. Once trained, data analysis by digital filtering can be performed quickly. Assessment of the reliability of the approach was performed through comparisons of simulated voltage transients, in which the ground truth results were known a priori. The LDA filter was also found to recover deconvolved impulses for single photon counting from highly distorted ringing waveforms from an impedance mismatched photomultiplier tube. The LDA filter was successful in removing these ringing distortions from two-photon excited fluorescence micrographs and through data simulations was found to extend the dynamic range of photon counting by approximately 3 orders of magnitude through minimization of detector paralysis.
منابع مشابه
Optical Filter Design of Fluorescence Imaging System Using Linear Disciriminant Analysis
Based on a fluorescence imaging model, the optical filters, such as excitation and emission filters, are determined by linear discriminant analysis. The discriminability is a quadratic Rayleigh quotient of one filter if the other filter is assumed to be constant. The optical filters are optimized alternatingly and iteratively by solving their generalized eigenvalue problems. The proposed method...
متن کاملQuantitative biorelevant profiling of material microstructure within 3D porous scaffolds via multiphoton fluorescence microscopy.
This study presents a novel approach, based on fluorescence multiphoton microscopy (MPM), to image and quantitatively characterize the microstructure and cell-substrate interactions within microporous scaffold substrates fabricated from synthetic biodegradable polymers. Using fluorescently dyed scaffolds fabricated from poly(DTE carbonate)/poly(DTO carbonate) blends of varying porosity and comp...
متن کاملMultiphoton fluorescence and second harmonic generation imaging of the structural alterations in keratoconus ex vivo.
PURPOSE To demonstrate the application of multiphoton fluorescence (MF) and second harmonic generation (SHG) microscopy for ex vivo characterization of the structural alterations of human corneas due to keratoconus. METHODS Buttons of keratoconic corneas derived from penetrating keratoplasty were sent for structural analysis with a custom-built multiphoton microscope. Fluorescence detected wi...
متن کاملMultiphoton excitation fluorescence microscopy of 5-aminolevulinic acid induced fluorescence in experimental gliomas.
BACKGROUND AND OBJECTIVE The clinical usefulness of 5-ALA guided detection of tumor tissue has been demonstrated for a number of malignancies. However, current techniques of intraoperative detection of protoporphyrin IX fluorescence in situ do not offer subcellular resolution. Therefore, discrimination of non-specific 5-ALA induced fluorescence remains difficult. MATERIALS AND METHODS In this...
متن کاملThe analysis of fluorophore orientation by multiphoton fluorescence microscopy
The accessibility of tunable, ultrafast laser sources has spurred the development and wide application of specialized microscopy techniques based on chromophore fluorescence following twoand three-photon absorption. The attendant advantages of such methods, which have led to a host of important applications including three-dimensional biological imaging, include some features that have as yet r...