Spectral and sequence analysis of finger blood pressure variability. Comparison with analysis of intra-arterial recordings.

نویسندگان

  • S Omboni
  • G Parati
  • A Frattola
  • E Mutti
  • M Di Rienzo
  • P Castiglioni
  • G Mancia
چکیده

The aim of our study was to assess whether the Finapres device is able to accurately monitor not only average blood pressure values but also blood pressure variability. To examine this issue, we analyzed 30-minute recordings of finger and intra-arterial pressure simultaneously obtained at rest in 14 patients. We compared systolic blood pressure, diastolic blood pressure, mean arterial pressure, pulse interval (the reciprocal of heart rate), overall variability (standard deviation), and specific time-domain and frequency-domain components. Systolic blood pressure, diastolic blood pressure, mean arterial pressure, and pulse interval spectral powers were computed by fast Fourier transform over three frequency bands: low frequency (0.025 to 0.07 Hz), midfrequency (0.07 to 0.14 Hz), and high frequency (0.14 to 0.35 Hz). The coherence, ie, the degree of association between blood pressure and pulse interval powers obtained by the two techniques, was also assessed. Standard deviations of diastolic blood pressure, mean arterial pressure, and pulse interval were similar when assessed from the two recordings, whereas standard deviation of systolic blood pressure was overestimated by analysis of finger pressure recordings. All powers of diastolic blood pressure and mean arterial pressure and high-frequency powers of systolic blood pressure estimated from analysis of finger blood pressure tracings were superimposable to those obtained by analyzing invasive recordings. Low-frequency and midfrequency powers of intra-arterial systolic blood pressure were significantly overestimated by the analysis of finger blood pressure tracings (+13.7 +/- 4.4 mm Hg2, P < .01, and +2.3 +/- 0.9 mm Hg2, P < .05).(ABSTRACT TRUNCATED AT 250 WORDS)

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Estimation of blood pressure variability from 24-hour ambulatory finger blood pressure.

Portapres is a noninvasive, beat-to-beat finger blood pressure (BP) monitor that has been shown to accurately estimate 24-hour intra-arterial BP at normal and high BPs. However, no information is available on the ability of this device to accurately track ambulatory BP variability. In 20 ambulatory normotensive and hypertensive subjects, we measured 24-hour BP by Portapres and through a brachia...

متن کامل

Comparison of finger and intra-arterial blood pressure monitoring at rest and during laboratory testing.

The accuracy of blood pressure values obtained by continuous noninvasive finger blood pressure recording via the FINAPRES device was evaluated by comparison with simultaneous intraarterial monitoring both at rest and during performance of tests known to induce fast and often marked changes in blood pressure. The comparison was performed in 24 normotensive or essential hypertensive subjects. The...

متن کامل

Anxiety and hostility are associated with reduced baroreflex sensitivity and increased beat-to-beat blood pressure variability.

OBJECTIVE The purpose of this study was to determine whether psychological factors are associated with heart rate variability (HRV), blood pressure variability (BPV), and baroreflex sensitivity (BRS) among healthy middle-aged men and women. METHODS A population-based sample of 71 men and 79 women (35-64 years of age) was studied. Five-minute supine recordings of ECG and beat-to-beat photoplet...

متن کامل

Effect of propofol on heart rate, arterial pressure and digital plethysmograph variability.

We have investigated the effects of propofol on beat-to-beat fluctuations in heart rate, arterial pressure (continuous arterial tonometry) and infrared finger plethysmography in 10 healthy male patients undergoing elective surgery. Administration of Diprivan (propofol 10 mg ml-1) 0.2 ml kg-1, followed by an infusion of 1 ml kg-1 h-1, was associated with transient tachycardia and a reduction in ...

متن کامل

Comparison of performance of various sphygmomanometers with intra-arterial blood-pressure readings.

Seven types of sphygmomanometer were used in random order on each of nine hypertensive patients and the readings compared with simultaneous intra-arterial blood-pressure recordings. All the devices gave significantly different values for systolic pressure, and only two measured diastolic pressure without significant error. Systolic pressure was consistently underestimated (range 31-7 mm Hg), an...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Hypertension

دوره 22 1  شماره 

صفحات  -

تاریخ انتشار 1993