Calibration of pneumotachographs using a calibrated syringe.

نویسندگان

  • Yongquan Tang
  • Martin J Turner
  • Johnny S Yem
  • A Barry Baker
چکیده

Pneumotachograph require frequent calibration. Constant-flow methods allow polynomial calibration curves to be derived but are time consuming. The iterative syringe stroke technique is moderately efficient but results in discontinuous conductance arrays. This study investigated the derivation of first-, second-, and third-order polynomial calibration curves from 6 to 50 strokes of a calibration syringe. We used multiple linear regression to derive first-, second-, and third-order polynomial coefficients from two sets of 6-50 syringe strokes. In part A, peak flows did not exceed the specified linear range of the pneumotachograph, whereas flows in part B peaked at 160% of the maximum linear range. Conductance arrays were derived from the same data sets by using a published algorithm. Volume errors of the calibration strokes and of separate sets of 70 validation strokes (part A) and 140 validation strokes (part B) were calculated by using the polynomials and conductance arrays. Second- and third-order polynomials derived from 10 calibration strokes achieved volume variability equal to or better than conductance arrays derived from 50 strokes. We found that evaluation of conductance arrays using the calibration syringe strokes yields falsely low volume variances. We conclude that accurate polynomial curves can be derived from as few as 10 syringe strokes, and the new polynomial calibration method is substantially more time efficient than previously published conductance methods.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improvement of Accuracy in Environmental Dosimetry by TLD Cards Using Three-dimensional Calibration Method

Background: The angular dependency of response for TLD cards may cause deviation from its true value on the results of environmental dosimetry, since TLDs may be exposed to radiation at different angles of incidence from the surrounding area.Objective: A 3D setting of TLD cards has been calibrated isotropically in a standard radiation field to evaluate the improvement of the accuracy of meas...

متن کامل

Calibration of iridium-192 source by ionization chamber for high dose rate brachytherapy

Background: The effectiveness and safety of brachytherapy treatment is mainly concerned with the calibration of sources and their traceability to internationally accepted Standards. Secondary Standard Dosimetry Laboratory (SSDL) does not offer calibration of ionization chambers with gamma-ray spectrum of high dose rate source 192Ir . This work has been carried out to calibrate the high dose rat...

متن کامل

Design and calibration of unicapillary pneumotachographs.

This study presents a method for design and calibration of unicapillary pneumotachographs for small-animal experiments. The design, based on Poiseuille's law, defines a set of internal radius and length values that allows for laminar flow, measurable pressure differences, and minimal interference with animal's respiratory mechanics and gas exchange. A third-order polynomial calibration (Pol) of...

متن کامل

Calibration of Area-Reduction Empirical Method for Dam Reservoir

The mathematical model of Area-Reduction empirical method is written through MATLAB software. Three periods of the hydrologic survey are required to run, calibrate and validatethe model. The Dez Dam, which is one of the most significant dams of the Middle East with three periods of hydrographic surveys in 1972, 2002 and 2011, has been chosen for the investigation. The annual average of the inpu...

متن کامل

Real Time Calibration of Strap-down Three-Axis-Magnetometer for Attitude Estimation

Three-axis-magnetometers (TAMs) are widely utilized as a key component of attitude determination subsystems and as such are considered the corner stone of navigation for low Earth orbiting (LEO) space systems. Precise geomagnetic-based navigation demands accurate calibration of the magnetometers. In this regard, a complete online calibration process of TAM is developed in the current research t...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of applied physiology

دوره 95 2  شماره 

صفحات  -

تاریخ انتشار 2003