Dynamic hip screw versus DHS blade: a biomechanical comparison of the fixation achieved by each implant in bone.
نویسندگان
چکیده
We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study.
منابع مشابه
Sliding hip screw versus sliding helical blade for intertrochanteric fractures: a propensity score-matched case control study.
The spiral blade modification of the Dynamic Hip Screw (DHS) was designed for superior biomechanical fixation in the osteoporotic femoral head. Our objective was to compare clinical outcomes and in particular the incidence of loss of fixation. In a series of 197 consecutive patients over the age of 50 years treated with DHS-blades (blades) and 242 patients treated with conventional DHS (screw) ...
متن کاملA Biomechanical Study Comparing Helical Blade with Screw Design for Sliding Hip Fixations of Unstable Intertrochanteric Fractures
Dynamic hip screw (DHS) is a well-established conventional implant for treating intertrochanteric fracture. However, revision surgery sometimes still occurs due to the cutting out of implants. A helical blade instead of threaded screw (DHS blade) was designed to improve the fixation power of the osteoporotic intertrochanteric fracture. In this study, the biomechanical properties of DHS blade co...
متن کاملComparison of Intertrochanteric Fracture Fixation with Dynamic Hip Screw and Bipolar Hemiarthroplasty Techniques
Background: At present, there is no consensus on an appropriate treatment modality for intertrochanteric fractures in the elderly with background diseases. The aim of the present study was to compare treatment outcomes of intertrochanteric fractures reduced with dynamic hip screws (DHS) and bipolar hemiarthroplasty in elderly patients with background medical conditions. Methods: In this randomi...
متن کاملMeasurement of Tip Apex Distance and Migration of Lag Screws and Novel Blade Screw Used for the Fixation of Intertrochanteric Fractures
Fixation with a dynamic hip screw (DHS) is one of the most common methods for stabilizing intertrochanteric fractures, except for unstable and reverse oblique fracture types. However, failure is often observed in osteoporotic patients whereby the lag screw effectively 'cuts out' through the weak bone. Novel anti-migration blades have been developed to be used in combination with a lag screw ('B...
متن کاملThe dynamic locking blade plate, a new implant for intracapsular hip fractures: biomechanical comparison with the sliding hip screw and Twin Hook.
Internal fixation of intracapsular hip fractures results in a high failure rate with non-union and avascular necrosis being the two most important complications. In order to prevent these possible complications treatment should consist of an anatomical reduction and stable fixation by insertion of a low volume, dynamic implant, providing angular and rotational stability to the femoral head. Acc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 93 5 شماره
صفحات -
تاریخ انتشار 2011