Towards Robot-Assisted Retinal Vein Cannulation: A Motorized Force-Sensing Microneedle Integrated with a Handheld Micromanipulator †
نویسندگان
چکیده
Retinal vein cannulation is a technically demanding surgical procedure where therapeutic agents are injected into the retinal veins to treat occlusions. The clinical feasibility of this approach has been largely limited by the technical challenges associated with performing the procedure. Among the challenges to successful vein cannulation are identifying the moment of venous puncture, achieving cannulation of the micro-vessel, and maintaining cannulation throughout drug delivery. Recent advances in medical robotics and sensing of tool-tissue interaction forces have the potential to address each of these challenges as well as to prevent tissue trauma, minimize complications, diminish surgeon effort, and ultimately promote successful retinal vein cannulation. In this paper, we develop an assistive system combining a handheld micromanipulator, called "Micron", with a force-sensing microneedle. Using this system, we examine two distinct methods of precisely detecting the instant of venous puncture. This is based on measured tool-tissue interaction forces and also the tracked position of the needle tip. In addition to the existing tremor canceling function of Micron, a new control method is implemented to actively compensate unintended movements of the operator, and to keep the cannulation device securely inside the vein following cannulation. To demonstrate the capabilities and performance of our uniquely upgraded system, we present a multi-user artificial phantom study with subjects from three different surgical skill levels. Results show that our puncture detection algorithm, when combined with the active positive holding feature enables sustained cannulation which is most evident in smaller veins. Notable is that the active holding function significantly attenuates tool motion in the vein, thereby reduces the trauma during cannulation.
منابع مشابه
Robotic Assisted Cannulation of Occluded Retinal Veins
PURPOSE To develop a methodology for cannulating porcine retinal venules using a robotic assistive arm after inducing a retinal vein occlusion using the photosensitizer rose bengal. METHODOLOGY Retinal vein occlusions proximal to the first vascular branch point were induced following intravenous injection of rose bengal by exposure to 532nm laser light delivered by slit-lamp or endolaser prob...
متن کاملEndovascular cannulation with a microneedle for central retinal vein occlusion.
We developed a new surgical treatment in which a microneedle is used for retinal endovascular cannulation to treat eyes with central retinal vein occlusion by flushing thrombus out of the central retinal vein as it passes through the lamina cribrosa. The eyes of 12 consecutive patients (12 eyes) with central retinal vein occlusion were successfully treated using this novel treatment. At 24 week...
متن کاملRelease of experimental retinal vein occlusions by direct intraluminal injection of ocriplasmin
PURPOSE Retinal vein occlusions (RVO) are a major cause of vision loss in people aged 50 years and older. Current therapeutic options limit the consequences of RVO but do not eliminate the cause. Cannulation of the involved vessel and removal of the clot may provide a more permanent solution with a less demanding follow-up. However, cannulation of smaller retinal veins remains challenging. This...
متن کاملPosition-Based Virtual Fixtures for Membrane Peeling with a Handheld Micromanipulator.
Peeling delicate retinal membranes, which are often less than 5 µm thick, is one of the most challenging retinal surgeries. Preventing rips and tears caused by tremor and excessive force can decrease injury and reduce the need for follow up surgeries. We propose the use of a fully handheld microsurgical robot to suppress tremor while enforcing helpful constraints on the motion of the tool. Usin...
متن کاملSemiautomated intraocular laser surgery using handheld instruments.
BACKGROUND AND OBJECTIVE In laser retinal photocoagulation, hundreds of dot-like burns are applied. We introduce a robot-assisted technique to enhance the accuracy and reduce the tedium of the procedure. MATERIALS AND METHODS Laser burn locations are overlaid on preoperative retinal images using common patterns such as grids. A stereo camera/monitor setup registers and displays the planned bu...
متن کامل