A Comparison of the Haider Tube-Guard® Endotracheal Tube Holder Versus Adhesive Tape to Determine if This Novel Device Can Reduce Endotracheal Tube Movement and Prevent Unplanned Extubation

نویسندگان

  • Jack C. Buckley
  • Adam P. Brown
  • John S. Shin
  • Kirsten M. Rogers
  • Nir N. Hoftman
چکیده

BACKGROUND Endotracheal tube security is a critical safety issue. We compared the mobility of an in situ endotracheal tube secured with adhesive tape to the one secured with a new commercially available purpose-designed endotracheal tube-holder device (Haider Tube-Guard). We also observed for the incidence of oropharyngeal or facial trauma associated with the 2 tube fixation methods. METHODS Thirty adult patients undergoing general anesthesia with neuromuscular blockade were prospectively enrolled. Immediately after intubation, a single study author positioned the endotracheal tube tip in the distal trachea using a bronchoscope. Anesthesiologists caring for patients secured the tube in their normal fashion (always with adhesive tape). A force transducer was used to apply linear force, increasing to 15 N or until the principal investigator deemed that the force be aborted for safety reasons. The displacement of the endotracheal tube was measured with the bronchoscope. Any tape was then removed and the endotracheal tube secured with the Haider Tube-Guard device. The linear force was reapplied and the displacement of the endotracheal tube measured. The Haider Tube-Guard device was left in place for the duration of the case. The patient's face and oropharynx were examined for any evidence of trauma during surgery and in the recovery room. On discharge from the postanesthesia care unit, the patient answered a brief survey assessing for any subjective evidence of minor facial or oropharyngeal trauma. RESULTS Under standardized tension, the endotracheal tube withdrew a mean distance of 3.4 cm when secured with adhesive tape versus 0.3 cm when secured with the Haider Tube-Guard (P <0.001). Ninety-seven percent of patients (29/30) experienced clinically significant endotracheal tube movement (>1 cm) when adhesive tape was used to secure the tube versus 3% (1/30) when the Haider Tube-Guard was used (P <0.001). Thirty percent of patients (9/30) were potentially deemed a high extubation risk (endotracheal tube movement >4 cm) when the endotracheal tube was secured with tape versus 0% (0/30) when secured with the Haider Tube-Guard (P = 0.004). Six patients with taped endotracheal tubes required the traction to be aborted before 15 N of force was achieved to prevent potential extubation as the tape either separated from the face or stretched to allow excessive endotracheal tube movement. None of the patients appeared to sustain any injury from the Haider Tube-Guard device. CONCLUSIONS The Haider Tube-Guard significantly reduced the mobility of the endotracheal tube when compared with adhesive tape and was well tolerated in our observations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Stabilizing Endotracheal Tube on Quality of Airway Management in Patients Hospitalized in ICU

Background: The unplanned departure of endotracheal tube is one of the main complications for endotracheal intubation. In addition to endotracheal extubation, its movement can also lead to damage to the oral cavity. Stabilizing endotracheal tube is the most important factor that can prevent unplanned departure of endotracheal tube as well as its movement. The current study was done with the aim...

متن کامل

Best Method for Securing an Endotracheal Tube.

CEN, CFRN, CPEN, and Tom Pietrantonio, BSRT, RRT-RCP, reply: In a study published in Respiratory Care, Shimizu et al3 concluded that the conventional adhesive tape methods had greater extubation force than the 2 endotracheal tube holders and that extubation force varies considerably depending on the width, length, and type of adhesive tape used. Mohammed and Hassan4 studied endotracheal tube sl...

متن کامل

Endotracheal tube extubation force: adhesive tape versus endotracheal tube holder.

BACKGROUND Adhesive tape is commonly used to secure the endotracheal tube (ETT) in anesthesia and intensive-care settings. OBJECTIVE To determine the force required to extubate when the ETT is secured with adhesive tape or commercially available ETT holders. METHODS We orally intubated a simulation manikin with a standard 8.0-mm inner-diameter ETT, inflated the cuff to 20 cm H(2)O, and mea...

متن کامل

Comparative assessment of the effects of securing endotracheal tube with endotracheal tube holder devices and the routine method on the quality of airway care in the intensive care unit

Background: Mechanical Ventilation in the intensive care unit is one of the most important interventions. Artificial airways are used to facilitate ventilation and endotracheal tubes are usually used for this matter. Unplanned extubation is one of the major complications of oral endotracheal intubation. In addition to extubation , tube movement can also cause trauma to oral cavity. Stabilizing ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 122  شماره 

صفحات  -

تاریخ انتشار 2016