Hydroxyapatite-coated femoral stems. Histology and histomorphometry around five components retrieved at post mortem.
نویسندگان
چکیده
We performed a histological and histomorphometric examination in five cadaver specimens of the femoral and acetabular components and the associated tissue which had been recovered between 3.3 and 6.2 years after primary total hip arthroplasty (THA) using a proximal hydroxyapatite (HA)-coated titanium alloy implant. All had functioned well during the patients' life. All the stems were fixed in the femur and showed osseointegration of both the proximal and distal parts. The amount of residual HA was greatest in the distal metaphyseal sections, indicating that the rate of bone remodelling may be the main factor causing loss of HA. The level of activity of the patient was the only clinical factor which correlated with loss of coating. The percentage of bone-implant osseointegration was almost constant, regardless of the amount of HA residue, periprosthetic bone density or the time of implantation. HA debris was seldom observed and if present did not cause any adverse or inflammatory reaction. Partial debonding did occur in one case as a result of a polyethylene-induced inflammatory reaction.
منابع مشابه
A comparison of bone remodelling around hydroxyapatite-coated, porous-coated and grit-blasted hip replacements retrieved at post-mortem.
We investigated the implant-bone interface around one design of femoral stem, proximally coated with either a plasma-sprayed porous coating (plain porous) or a hydroxyapatite porous coating (porous HA), or which had been grit-blasted (Interlok). Of 165 patients implanted with a Bimetric hip hemiarthroplasty (Biomet, Bridgend, UK) specimens were retrieved from 58 at post-mortem. We estimated ing...
متن کاملBone lysis in well-fixed cemented femoral components.
We have reviewed 25 cases of focal femoral osteolysis in radiographically stable, cemented femoral implants. In three hips retrieved at post-mortem from two patients, we have been able to make a detailed biomechanical and histological analysis. The interval between arthroplasty and the appearance of focal osteolysis on clinical radiographs ranged from 40 to 168 months, and in over 70% of the ca...
متن کاملThe Effect of Hydroxyapatite Coating on Long-term Results of Total Hip Arthroplasty with Hydroxyapatite-coated Anatomic Femoral Stem
PURPOSE To evaluate the clinical and radiological results, as well as the survival rate, associated with total hip arthroplasty using a hydroxyapatite (HA)-coated anatomical femoral stem at a follow-up of ≥12 years. MATERIALS AND METHODS From April 1992 to May 1997, 86 patients (102 hips) underwent total hip arthroplasty with a HA-coated ABG I (Anatomical Benoist Giraud; Howmedica) hip prosth...
متن کاملFocal femoral osteolysis after revision hip replacement with a cannulated, hydroxyapatite-coated long-stemmed femoral component: a new route for particulate wear debris.
We describe a case of symptomatic focal femoral osteolysis around a screw hole distal to the hydroxyapatite-coated portion of a cannulated femoral component in a revision hip replacement. No locking screw had been inserted into this, the most proximal of the three distal holes for locking screws. The presence of polyethylene wear debris in the tissue excised from the lesion suggested that it ha...
متن کاملHydroxyapatite coating on carbon composite hip implants in dogs.
In 33 dogs we implanted femoral stems made either of carbon composite, some coated with hydroxyapatite, or of titanium alloy with a porous coating. Osseointegration was greater in the hydroxyapatite-coated than in the un-coated stems (p less than 0.001). Push-out tests, at an average of 7.2 months after implantation, showed a six-fold increase in interface shear strength and a twelve-fold incre...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 81 1 شماره
صفحات -
تاریخ انتشار 1999