Reproducibility of positive test results in the BDProbeTec ET system for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

نویسندگان

  • Elizabeth E Culler
  • Angela M Caliendo
  • Frederick S Nolte
چکیده

Nucleic acid amplification tests such as the BDProbeTec ET (BDPT) system are more prone to reproducibility problems than are antigen detection tests and culture. A repeat testing algorithm for all samples with method other than acceleration (MOTA) scores greater than or equal to the cutoff value (2000) was developed for the BDPT system and applied in a clinical laboratory setting. All positive samples were retested, and if the result of the second test was below the cutoff value, a third test was performed to resolve the discrepancy. Overall, 11 (5.3%) of 207 samples initially positive for Chlamydia trachomatis and 11 (10.7%) of 103 samples initially positive for Neisseria gonorrhoeae were not confirmed by repeat testing of the original sample. Poor reproducibility was associated with low-positive MOTA scores (2000 to 9999) for both analytes. Only 21 (80.8%) of 26 low-positive samples in the C. trachomatis test and 4 (33.3%) of 12 low-positive samples in the N. gonorrhoeae test retested as positive. The reproducibility of both tests with samples with initial MOTA scores of >or=10000 increased to 96.7%. The data suggest that retesting of low-positive samples is warranted and could reduce the number of potentially false-positive test results.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Multicenter evaluation of the BDProbeTec ET System for detection of Chlamydia trachomatis and Neisseria gonorrhoeae in urine specimens, female endocervical swabs, and male urethral swabs.

The performance of the Becton Dickinson BDProbe Tec ET System Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays (BD Biosciences, Sparks, Md.) was evaluated in a multicenter study. Specimens were collected from 2,109 men and women, with or without symptoms, attending sexually transmitted disease, family planning, and obstetrics and gynecology clinics. Both swab and urine sampl...

متن کامل

شناسایی نایسریا گونورآ و کلامیدیا تراکوماتیس در بیماران مبتلا به اورتریت علامت دار با استفاده از روش های Multiplex PCR، رنگ آمیزی گرم و کشت ادرار در یاسوج

Background and purpose: Chlamydia trachomatis and Neisseria gonorrhoeae are the most common causes of sexually transmitted infections. The aim of this study was detection of N. gonorrhoeae and C. trachomatis in patients with symptomatic urethritis using Multiplex PCR, gram stain and urine culture. Materials and methods: This cross sectional study was conducted in 137 patients with symptomatic ...

متن کامل

Direct comparison of the BD ProbeTec ET system with in-house LightCycler PCR assays for detection of Chlamydia trachomatis and Neisseria gonorrhoeae from clinical specimens.

The commercial BD ProbeTec ET (BDPT) system for the detection of Neisseria gonorrhoeae and Chlamydia trachomatis from clinical specimens was compared with our in-house LightCycler real-time PCR (LC-PCR) assays. Specimens initially positive by the BDPT system were retested by our LC-PCR assays. Our results for C. trachomatis testing indicate a 91.2% agreement when the results of 114 clinical spe...

متن کامل

Characteristics of the m2000 automated sample preparation and multiplex real-time PCR system for detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

We evaluated a new real-time PCR-based prototype assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae developed by Abbott Molecular Inc. This assay is designed to be performed on an Abbott m2000 real-time instrument system, which consists of an m2000sp instrument for sample preparation and an m2000rt instrument for real-time PCR amplification and detection. The limit of de...

متن کامل

Performance of strand displacement amplification assay in the detection of Chlamydia trachomatis and Neisseria gonorrhoeae.

This study is a critical analysis of certain amplification assays for detecting Chlamydia trachomatis and Neisseria gonorrhoeae infections which have demonstrated that the plasmid-free variant of C. trachomatis is frequently responsible for infection in our patients. Specifically, we evaluated the performance of the strand displacement amplification (SDA) assay in detecting either C. trachomati...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of clinical microbiology

دوره 41 8  شماره 

صفحات  -

تاریخ انتشار 2003