A Clinical Trial of the Accuracy and Treatment Experience of the Flash Glucose Monitor FreeStyle Libre in Adults with Type 1 Diabetes
نویسندگان
چکیده
BACKGROUND In Sweden, FreeStyle Libre a flash glucose monitoring system came onto the market in 2014 as a complement to self-monitoring of blood glucose. The aim of this study was to evaluate the accuracy and treatment experience of the FreeStyle Libre system. METHODS Fifty-eight adults with type 1 diabetes used FreeStyle Libre for 10-14 days and measured capillary blood glucose levels with the HemoCue blood glucose measurement system at least six times a day simultaneously. RESULTS For the entire study period, the mean absolute relative difference (MARD) was 13.2% (95% confidence interval [CI] 12.0%-14.4%). MARD was 13.6% (95% CI 12.1%-15.4%) during week 1 and 12.7% (95% CI 11.5%-13.9%) during week 2. The mean absolute difference (MAD) for the whole study period was 19.8 mg/dL (1.1 mmol/L) (95% CI 17.8-21.8 mg/dL), including 20.5 mg/dL (1.14 mmol/L) during week 1 and 19.0 mg/dL (1.05 mmol/L) during week 2. The overall correlation coefficient was 0.96. For glucose values <72, 72-180, and >180 mg/dL (<4, 4-10, and >10 mmol/L), the MARD was 20.3% (95% CI 17.7%-23.1%), 14.7% (95% CI 13.4%-16%), and 9.6% (95% CI 8.5%-10.8%), respectively, and respective MAD values were 12.3, 17.8, and 23.6 mg/dL (0.69, 0.99, and 1.31 mmol/L). Using the 10-item visual analog scale, patients rated their experience with FreeStyle Libre as generally positive, with mean values ranging from 8.22 to 9.8. CONCLUSIONS FreeStyle Libre had a similar overall MARD as continuous blood glucose monitoring systems in earlier studies when studied in similar at-home conditions. The overall patient satisfaction was high.
منابع مشابه
Accuracy, User Acceptability, and Safety Evaluation for the FreeStyle Libre Flash Glucose Monitoring System When Used by Pregnant Women with Diabetes
BACKGROUND Accuracy of the FreeStyle Libre™ Flash Glucose Monitoring System has not been evaluated in pregnant women with diabetes. The aim of this study was to determine accuracy (compared to self-monitoring of blood glucose [SMBG]), clinical safety, and acceptability of the FreeStyle Libre System when used at home by this population. MATERIALS AND METHODS Seventy-four participants, with typ...
متن کاملThe performance of flash glucose monitoring in critically ill patients with diabetes.
OBJECTIVE Frequent glucose monitoring may improve glycaemic control in critically ill patients with diabetes. We aimed to assess the accuracy of a novel subcutaneous flash glucose monitor (FreeStyle Libre [Abbott Diabetes Care]) in these patients. METHODS We applied the FreeStyle Libre sensor to the upper arm of eight patients with diabetes in the intensive care unit and obtained hourly flash...
متن کاملThe Performance and Usability of a Factory-Calibrated Flash Glucose Monitoring System
INTRODUCTION The purpose of the study was to evaluate the performance and usability of the FreeStyle(®) Libre™ Flash glucose monitoring system (Abbott Diabetes Care, Alameda, CA) for interstitial glucose results compared with capillary blood glucose results. MATERIALS AND METHODS Seventy-two study participants with type 1 or type 2 diabetes were enrolled by four U.S. clinical sites. A sensor ...
متن کاملClinical Use of Continuous Glucose Monitoring in Adults with Type 1 Diabetes
With the emphasis on intensive management of type 1 diabetes, data from studies support frequent monitoring of glucose levels to improve glycemic control and reduce glucose variability, which can be related to an increase in macro and microvascular complications. However, few perform capillary blood glucose that frequently. There are currently two available alternatives that this review will di...
متن کاملFlash Glucose Monitoring: The Future Is Here
We cannot believe that it has been more than two decades since we first evaluated real-time continuous glucose monitoring (CGM)—(GlucoWatch ) in patients with type 1 diabetes (T1D). The first real-time CGM (GlucoWatch), which uses a reversed iontophoresis to measure interstitial glucose, was approved to be used for 12 h by the Food and Drug Administration (FDA) in 1999. Since then, several adva...
متن کامل