Extended clinical trial and evaluation of glucose determination with the Eastman Kodak Ektachem GLU/BUN Analyzer.
نویسندگان
چکیده
We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analyses for glucose with the "Ektachem." We analyzed 760 clinical samples by this technique, by the FDA Proposed Class Standard glucose reference method, and with the Beckman System I GLU/BUN Analyzer. Precision and accuracy were estimated for 500, 1000, 1200, 1500, and 3000 mg/L glucose concentrations in 100, 30, or 20 microL of serum or plasma. Potential interference of 19 compounds was evaluated. Random error (1.965 X SD) was 22, 30, 34, 40, and 88 mg/L. Systematic error was 8, 1.5, -2, -5, and -27 mg/L. Total analytical error was 30, 32, 36, 46, and 110 mg/L for analysis of 100 microL of serum at the above-stated glucose concentrations. The greatest interference (-39 mg/L) in the glucose (at 1200 mg/L) determination was caused by L-ascorbate (40 mg/L). Glucose concentrations as determined with the Ektachem were found to be linearly related to the expected concentration up to at least 5660 mg/L. Carryover was not statistically significant.
منابع مشابه
Extended clinical trial and evaluation of urea nitrogen determination with the Ektachem GLU/BUN analyzer.
We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analysis for urea nitrogen with the multilayer film analysis system ("Ektachem"). We analyzed 456 clinical samples with this instrument, by the manual urease/glucose dehydrogenase method, and with the Beckman System I GLU/BUN An...
متن کاملAbstracts/Sommaires/Zusammenfassungen
An evaluation of the Kodak glucose/BUN analyzer including experience with proposed testing protocols An evaluation of the analysis of serum urea and plasma_glucose on the Kodak Ektachem GLU/BUN()analyzer is described. It incorporates the experimental design and statistical methods outlined in the Proposed Standard for Establishing Performance Claims (PSEP) (1) for clinical chemical methods PSEP...
متن کاملEvaluation of an engineering model of the "EKTACHEM" analyzer for glucose and urea assay.
We report results of a 21-week evaluation of an engineering model of the Kodak “EKTACHEM Glu/BUN Analyzer.” This novel clinical instrument measures glucose and “urea nitrogen” (actually, urea) in a minimum of 30 j.tL of body fluids by use of the principles of thin, multilayer film analysis and reflectance spectrophotometry. We compared the performance of the instrument to that of the Dupont aca...
متن کاملClinical evaluation of Eastman Kodak's Ektachem 400 Analyzer.
We evaluated the performance of the Kodak Ektachem 400 Analyzer in a 16-week clinical trial. We assessed four potentiometric tests and nine colorimetric tests for precision and correlation with results obtained some other commonly used instruments (Technicon SMA II and C800 System, Du Pont aca II, and Baker CentrifiChem). The comparison was favorable for all tests except albumin, sodium, and ca...
متن کاملThe determination of calcium, glucose, urea and uric acid using the Kodak EKTACHEM multilayer film technology: an evaluation.
Calcium, glucose, urea and uric acid determinations on the "'Kodak Ektachem Four Chemistry Analyzer" were tested for precision and an estimation of the accuracy was performed with our laboratory routine methods using patients' sera and proficiency fluids. Precision -- examined over a period of six weeks -- was very good. Even the very stringent claim of the College of American Pathologists (CAP...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 27 1 شماره
صفحات -
تاریخ انتشار 1981