Low hematocrit and serum albumin concentrations underlie the overestimation of tacrolimus concentrations by microparticle enzyme immunoassay versus liquid chromatography-tandem mass spectrometry.
نویسندگان
چکیده
BACKGROUND Rapid liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods are used increasingly for tacrolimus (TRL) monitoring but show a negative difference with respect to a microparticle immunoassay (MEIA). This report examines possible reasons for this difference between methods. METHODS We collected 1156 blood samples from 277 adult and 121 pediatric recipients of liver, renal, and bone marrow grafts or hepatocyte or pancreatic islet cell implants. TRL was measured in whole blood by MEIA and LC-MS/MS, and hematologic and biochemical data were collected when available. RESULTS LC-MS/MS was significantly more precise (P <0.02) than the MEIA with increased sensitivity. The MEIA had a median difference of 16.2% vs LC-MS/MS overall, and this was significantly affected by patient cohort (P <0.001). The difference was greater in adult or pediatric liver graft recipients while they were inpatients rather than outpatients (31.8% and 14.0% vs 7.5% and 6.5%, respectively). The difference was also greater in bone marrow than kidney graft recipients (32.8% vs 15.8%, respectively). Multiple linear regression analysis showed significant inverse relationships of this difference with hematocrit (packed cell volume) and plasma albumin (P <0.001) in the total cohort and a positive relationship with plasma bilirubin in a subgroup of pediatric liver graft recipients. CONCLUSIONS Patients with a low packed cell volume and plasma albumin are likely to show artificially high concentrations of TRL when measured by MEIA. The increased risk of underimmunosuppression must be considered should doses be reduced to lower these seemingly high TRL concentrations.
منابع مشابه
Evaluation of the tacrolimus II microparticle enzyme immunoassay (MEIA II) in liver and renal transplant recipients.
We evaluated the MEIA II with blood samples with added tacrolimus (3.0, 5.0, 11.0, and 22.0 microg/L). The assay had acceptable recoveries (99-103%) and intraday imprecision (<16.0%) across the range of concentrations studied, except for the recoveries at 3.0 microg/L (86.3%) and 5.0 microg/L (80.7%). Comparison of liver (n = 116) and renal (n = 113) patient samples measured by MEIA II against ...
متن کاملTacrolimus measurement: building a better immunoassay.
Tacrolimus (formerly known as FK506) is a macrolide immunosuppressant that is used for primary immunosuppression after organ transplantation (1 ). The landmark work of Starzl at the University of Pittsburgh paved the way for the clinical use of tacrolimus as a primary immunosuppressant, leading to a shift away from cyclosporine use (2– 4 ). Today, tacrolimus is used in the majority of transplan...
متن کاملMulti-site analytical evaluation of the Abbott ARCHITECT tacrolimus assay.
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Tacrolimus immunoassay. Proficiency panels and specimens from a population of organ transplant recipients were analyzed in 6 clinical laboratories in Europe and the United States, and the results were compared with other methods. The ARCHITECT assay requires a whole blood specimen pretreatment step wi...
متن کاملTime of drug administration, CYP3A5 and ABCB1 genotypes, and analytical method influence tacrolimus pharmacokinetics: a population pharmacokinetic study.
Tacrolimus (TAC) pharmacokinetics are characterized by a very high variability that complicates its therapeutic use. The aims of this study were: 1) to identify and model the effect of demographic, clinical, and genetic factors and time of drug administration on TAC pharmacokinetic variability; and 2) to assess the influence of the analytical method by modeling the TAC blood concentrations meas...
متن کاملHPLC-microparticle enzyme immunoassay specific for tacrolimus in whole blood of hepatic and renal transplant patients.
Tacrolimus is a relatively new immunosuppressant used in organ transplantation to prevent graft rejection. However, its use is not devoid of side effects, making it important to maintain blood concentrations within therapeutic ranges. Several analytical methods are currently available for routine drug monitoring. However, these methods are based on use of the same monoclonal antibody, which als...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 51 3 شماره
صفحات -
تاریخ انتشار 2005