Maximum dislodging forces of implant overdenture stud attachments.

نویسندگان

  • Vicki C Petropoulos
  • Woollcott Smith
چکیده

PURPOSE The aim of this study was to compare the retention and stability of the Nobel Biocare standard ball (NBS), Nobel Biocare 2.25-mm-diameter ball (NB2), Zest Anchor (ZA), Zest Anchor Advanced Generation (ZAAG), Sterngold ERA orange (SEO), and Sterngold ERA white (SEW) attachments on an implant-retained overdenture model. MATERIALS AND METHODS The attachments were tested using 2 permanently placed Brånemark System implants on a test model that was attached to an Instron machine (crosshead speed 50.8 mm/minute). Each attachment had one part embedded in a denture-like housing and the other part screwed into the implants. Dislodging tensile forces were applied to the housings in 3 directions simulating function: vertical, oblique, and anterior-posterior. Eight tests were done in 3 directions with 6 samples of each attachment. The dislodging forces generated measurements of the peak load (maximum dislodging force). A 1-way analysis of variance followed by the Tukey studentized range test was used to determine groups that were significantly different. All tests for significance were carried out at the .05 level of significance. RESULTS Results showed the ZAAG attachment to be the most retentive for the peak load measurement when subjected to vertically directed forces, with mean values and standard deviations of 37.2 +/- 5.5 N. The next most retentive attachment was the NBS, followed by the SEO, NB2, SEW, and ZA. For obliquely directed forces, the ZAAG attachment was the most retentive, with mean values and standard deviations of 27.2 +/- 4.2 N. The next most retentive was the NBS, followed by the NB2, SEO, ZA, and SEW. For anterior-posteriorly directed forces, results showed the NBS had the highest measured retentive force, with mean values and standard deviations of 34.6 +/- 18.8 N, but this was not statistically different from the NB2 and ZAAG; this was followed by the SEO, SEW, and ZA. DISCUSSION There has been a marked resurgence in the treatment of patients with overdentures using implant attachments as retentive devices. The maximum force developed (a measure of retention) as the implant stud attachments were resisting removal from the implant abutments was determined. CONCLUSIONS Based on the present study, the clinician may be able to make empirical decisions on attachment selection, depending on the amount of retention desired and the specific clinical situation.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of the Effect of Ball and Bar Attachment Designs on Retention and Stability of Mandibular Implant-Supported Overdentures

Objectives: The present study evaluated the effect of bar and ball attachment designs on retention and stability of a mandibular overdenture supported by four implants. Methods: An edentulous mandibular acrylic resin model with four implants in the anterior part of the ridge (A, B, D and E) was fabricated. A metal framework simulating the overdenture was also fabricated. Totally, 30...

متن کامل

Comparison of Three Attachments in Implant-Tissue Supported Overdentures: An In Vitro Study

OBJECTIVE Retention and resistance of the implant-tissue supported overdenture may be affected by the type of attachment. The aim of this research was to compare the retention and resistance of Nobel Biocare Ball (NBB), Nobel Biocare Bar and Clip (NBBC) and Sterngold ERA Red (ERAR) attachments on an implant-tissue supported overdenture model. MATERIALS AND METHODS The attachment samples were ...

متن کامل

The Effect of Inter-Implant Distance on Retention and Resistance to Dislodging Forces for Mandibular Implant-Tissue-Supported Overdentures

OBJECTIVES The effect of inter-implant distance on retention and resistance of implant-tissue-supported overdentures is lacking in the literature. An in vitro study was performed to evaluate this effect for mandibular implant-tissue-supported overdentures retained by two ball attachments. MATERIALS AND METHODS An acrylic cast of an edentulous mandible was fabricated. Three pairs of implants w...

متن کامل

Retention of maxillary implant overdenture bars of different designs.

STATEMENT OF PROBLEM The specific degree of retention for overdenture attachments is unknown in relation to design, location, and alignment to supporting dental implants. PURPOSE The purpose of this study was to evaluate the initial retention characteristics of 5 implant maxillary overdenture designs under in vitro dislodging forces. MATERIAL AND METHODS A simulated edentulous maxilla was f...

متن کامل

The influence of the number and the type of magnetic attachment on the retention of mandibular mini implant overdenture

PURPOSE The aim of this study was to compare the retention of mini implant overdenture by the number, the type of magnetic attachment, and the directions of applied dislodging force. MATERIALS AND METHODS The experimental groups were designed by the number and type of magnetic attachment. Twenty samples were tested with Magden implants. Each attachment was composed of the magnet assembly in o...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The International journal of oral & maxillofacial implants

دوره 17 4  شماره 

صفحات  -

تاریخ انتشار 2002