Prospective randomised comparison of irrigated-tip and large-tip catheter ablation of cavotricuspid isthmus-dependent atrial flutter.

نویسندگان

  • Christophe Scavée
  • Pierre Jaïs
  • Li-Fern Hsu
  • Prashanthan Sanders
  • Meleze Hocini
  • Rukshen Weerasooriya
  • Laurent Macle
  • Florence Raybaud
  • Jacques Clementy
  • Michel Haïssaguerre
چکیده

BACKGROUND Radiofrequency (RF) ablation of cavotricuspid isthmus (CTI) dependent flutter can be performed using different types of ablation catheters. It has been proposed that irrigated and large-tip catheters are capable of creating larger lesions, resulting in greater efficacy. This prospective, randomised clinical study compared the efficacy of irrigated and large-tip catheters of different designs. METHODS Eighty patients (69 men, 66+/-11 years) undergoing de novo RF ablation of CTI-dependent flutter were randomised to ablation using one of the following catheters: (i) externally-irrigated 20), (ii) internally-cooled (n=20), (iii) single sensor, 8-mm tip (n=20), or (iv) double sensor, 8-mm tip (n=20). The study endpoint was the demonstration of bidirectional CTI conduction block within 12 min of cumulative RF delivery. Crossover to the externally-irrigated catheter was permitted if this was not achieved. The ablation and procedural parameters, safety and efficacy were compared. RESULTS The primary endpoint was achieved in 64 patients (80%), including all 20 patients randomised to the externally-irrigated catheter. Crossover was required in 16 patients: 9 initially using the internally-cooled catheter (45%), 3 using single-sensor, 8-mm-tip (15%), and 4 using double-sensor, 8-mm-tip (20%) catheters. The higher initial failure rate with the internally-cooled-tip catheter was significant compared to the externally-irrigated (p = 0.001) and single-sensor, 8-mm-tip (p = 0.04) catheters. The externally-irrigated catheter achieved the study endpoint more frequently with fewer RF applications of shorter duration compared to the internally-cooled-tip catheter and 8-mm-tip catheters, the difference being significant compared with internally cooled ablation. No major complications were observed. CONCLUSION Among commonly used ablation catheters, the externally-irrigated catheter has a higher efficacy for rapid achievement of CTI block.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Successful irrigated-tip catheter ablation of atrial flutter resistant to conventional radiofrequency ablation.

BACKGROUND Catheter ablation of typical right atrial flutter is now widely performed. The best end point has been demonstrated to be bidirectional isthmus block. We investigated the use of irrigated-tip catheters in a small subset of patients who failed isthmus ablation with conventional radiofrequency (RF) ablation. METHODS AND RESULTS Of 170 patients referred for ablation of common atrial f...

متن کامل

Comparison of 8-mm-tip and irrigated-tip catheters in the ablation of isthmus-dependent atrial flutter: a prospective randomized trial.

INTRODUCTION AND OBJECTIVES Both 8-mm-tip and irrigated-tip catheters improve outcomes in typical atrial flutter ablation. However, it is not yet known which is preferable. The objective was to compare the effectiveness of 8-mm-tip (Group 1) and open irrigated-tip (Group 2) catheters in the first ablation attempt. METHODS A prospective randomized trial with a long-term follow-up was performed...

متن کامل

Prospective randomized comparison of irrigated-tip versus conventional-tip catheters for ablation of common flutter.

BACKGROUND Radiofrequency (RF) ablation of common flutter requires the creation of a complete ablation line to produce bidirectional conduction block in the cavotricuspid isthmus. An irrigated-tip ablation catheter has been shown to be effective in patients in whom conventional ablation has failed. This randomized study compares the efficacy and safety of this catheter with those of a conventio...

متن کامل

[Lesions produced by radiofrequency ablation of the cavotricuspid isthmus in an experimental model].

INTRODUCTION AND OBJECTIVES Experimental studies have shown that deeper and wider lesions (up to 10 mm long or deep) can be safely created using an 8 mm or irrigated tip catheter for ablation to treat atrial flutter. However, potential damage to the tricuspid valve or inferior cava vein has not been systematically evaluated. PATIENTS AND METHOD The cavotricuspid isthmus was ablated in 26 pigs...

متن کامل

Radiofrequency catheter ablation of common-type atrial flutter guided by conventional versus electroanatomical mapping.

Radiofrequency catheter (RF) ablation aiming the complete conduction block of the cavotricuspid isthmus has become treatment of choice for common type atrial flutter. Different approaches to guide the ablation procedure are used. For the conventional approach multipolar catheters in the right atrium and in the coronary sinus are required to detect the induction of conduction block via the isthm...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European heart journal

دوره 25 11  شماره 

صفحات  -

تاریخ انتشار 2004