Cataract surgery causes biomechanical alterations to the eye detectable by Corvis ST tonometry

نویسندگان

  • Yoshitake Kato
  • Shunsuke Nakakura
  • Ryo Asaoka
  • Kanae Matsuya
  • Yuki Fujio
  • Yoshiaki Kiuchi
چکیده

PURPOSE Modern cataract surgery is generally considered to bring about modest and sustained intraocular pressure (IOP) reduction. However, the pathophysiological mechanism for this remains unclear. Moreover, a change in ocular biomechanical properties after surgery can affect the measurement of IOP. The aim of the study is to investigate ocular biomechanics, before and following cataract surgery, using Corvis ST tonometry (CST). PATIENTS AND METHODS Fifty-nine eyes of 59 patients with cataract were analyzed. IOP with Goldmann applanation tonometry (IOP-G), axial length, corneal curvature and CST parameters were measured before cataract surgery and, up to 3 months, following surgery. Since CST parameters are closely related to IOP-G, linear modeling was carried out to investigate whether there was a change in CST measurements following cataract surgery, adjusted for a change in IOP-G. RESULTS IOP-G significantly decreased after surgery (mean±standard deviation: 11.8±3.1 mmHg) compared to pre-surgery (15.2±4.3 mmHg, P<0.001). Peak distance (the distance between the two surrounding peaks of the cornea at the highest concavity), maximum deformation amplitude (the movement of the corneal apex from the start of deformation to the highest concavity) and A1/A2 velocity (the corneal velocity during inward or outward movement) significantly increased after cataract surgery (P<0.05) while radius (the central curvature radius at the highest concavity) was significantly smaller following cataract surgery (P<0.05). Linear modeling supported many of these findings, suggesting that peak distance, maximum deformation amplitude and A2 velocity were increased, whereas A2 deformation amplitude and highest concavity time were decreased (after adjustment for IOP change), following cataract surgery. CONCLUSION Corneal biomechanical properties, as measured with CST, were observed to change significantly following cataract surgery. TRIAL REGISTRATION Japan Clinical Trials Registry UMIN000014370.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Erratum: A cross-sectional study to compare intraocular pressure measurement by sequential use of Goldman applanation tonometry, dynamic contour tonometry, ocular response analyzer, and Corvis ST

OBJECTIVE To study the correlation and effect of sequential measurement of intraocular pressure (IOP) with Goldmann applanation tonometer (GAT), ocular response analyzer (ORA), dynamic contour tonometer (DCT), and Corvis ST. SETTING AND DESIGN Observational cross-sectional series from the comprehensive clinic of a tertiary eye care center seen during December 2012. METHODS One hundred and t...

متن کامل

Biomechanical Characterization of Subclinical Keratoconus Without Topographic or Tomographic Abnormalities.

PURPOSE To present a case series of patients with subclinical keratoconus with normal topometric (anterior curvature) and tomographic findings in one eye who showed abnormalities detected by Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) in vivo biomechanical assessment. METHODS All patients had a complete ophthalmic examination, including the Corvis ST biomechanical measurements, opti...

متن کامل

Effects of age on corneal deformation by non-contact tonometry integrated with an ultra-high-speed (UHS) Scheimpflug camera.

PURPOSE To correlate parameters derived from corneal deformation resulting from non-contact tonometry integrated with an ultra-high-speed (UHS) Scheimpflug camera (Oculus Corvis ST, Scheimpflug Technology; Wetzlar, Germany) with age in normal eyes from young patients. METHODS Observational, retrospective study involving one eye randomly selected from study participants, totaling 89 healthy ey...

متن کامل

Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea.

PURPOSE The aim of this study was to show the usefulness of three methods for measuring IOP: Goldmann applanation tonometry, rebound tonometry, and Ultra-High-Speed Scheimpflug technology. METHODS The examined group consisted of 96 patients (192 eyes), including 63 women and 33 men with a mean age of 59.3 ± 19.9 years. Together, 152 healthy eyes and 40 eyes with different pathologies were exa...

متن کامل

Corneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas.

AIM To analyze the correlations between ocular biomechanical and biometric data of the eye, measured by Scheimpflug-based devices on healthy subjects. METHODS Three consecutive measurements were carried out using the corneal visualization Scheimpflug technology (CorVis ST) device on healthy eyes and the 10 device-specific parameters were recorded. Pentacam HR-derived parameters (corneal curva...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 12  شماره 

صفحات  -

تاریخ انتشار 2017