Performance of Abusign drugs-of-abuse slide tests with particluar emphasis on concentrations near the cutoff: comparison with FPIA-ADx and confirmation of results with GC-MS.

نویسندگان

  • J J Ros
  • M G Pelders
  • A C Egberts
چکیده

The objective of the present study was to evaluate the performance of Abusign test slides in comparison with fluorescence polarization immunoassay (FPIA)-ADx and gas chromatography-mass spectrometry (GC-MS) confirmation with special emphasis on concentrations near the cutoff. Analysis was performed on 35 individual slide tests for cocaine (cutoff, 300 ng/mL), cannabinoids (cutoff, 100 ng/mL), and opiates (cutoff, 300 ng/mL); 104 tests for cannabinoids (cutoff, 50 ng/mL); and 34 panel slides (amphetamine cutoff 1000 ng/mL, cocaine cutoff 300 ng/mL, opiates cutoff 300 ng/mL, and cannabinoids cutoff 100 ng/mL). Urine samples that had a concentration measured with FPIA-ADx within +/-25% of the Abusign cutoff were selected. Inter- and intra-individual agreement were assessed from the readings of the slides (positive or negative) by four persons at t = 3, 5, and 10 min after sample incubation. In addition, the sensitivity and specificity of the cannabinoids slide test (50 ng/mL) were compared with those of FPIA-ADx using GC-MS as the gold standard. There was a considerable variation in the reading of the slides between different persons. In addition, for the same observer the reading could depend upon the time of incubation. In comparison with FPIA-ADx, the Abusign test slides showed a high sensitivity (46% vs. 87%) and a low specificity (95% vs. 51%). We concluded that the Abusign slide test is unsuitable for situations in which a reliable test result is desired because of the low interindividual agreement, the dependency of the test result upon time, and the low specificity. However, the test slides may be of value in toxicological screening.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Oral fluid testing for drugs of abuse: positive prevalence rates by Intercept immunoassay screening and GC-MS-MS confirmation and suggested cutoff concentrations.

Draft guidelines for the use of oral fluid for workplace drug testing are under development by the Substance Abuse and Mental Health Services Administration (SAMHSA) in cooperation with industry and researchers. Comparison studies of the effectiveness of oral fluid testing versus urine testing are needed to establish scientifically reliable cutoff concentrations for oral fluid testing. We prese...

متن کامل

Oral Fluid Testing for Drugs of Abuse: Positive Prevalence Rates by Intercepff Immunoassay Screening and GC-MS-MS Confirmation and Suggested Cutoff Concentrations

Draft guidelines for the use of oral fluid for workplace drug testing are under development by the Substance Abuse and Mental Health Services Administration (SAMHSA) in cooperation with industry and researchers. Comparison studies of the effectiveness of oral fluid testing versus urine testing are needed to establish scientifically reliable cutoff concentrations for oral fluid testing. We prese...

متن کامل

Sensitivity, specificity, and efficiency in detecting opiates in oral fluid with the Cozart Opiate Microplate EIA and GC-MS following controlled codeine administration.

Oral fluid specimens (N = 1406) were collected from 19 subjects prior to and up to 72 h following controlled administration of oral codeine. Volunteers provided informed consent to participate in this National Institute on Drug Abuse Institutional Review Board-approved protocol. A modification of Cozart Microplate Opiate EIA Oral Fluid Kit (Opiate ELISA), employing codeine calibrators, was used...

متن کامل

Multianalyte assay system developed for drugs of abuse.

A simple, 10-min immunoassay system has been developed that simultaneously screens for five different classes of drugs of abuse in a urine sample. This system tests for amphetamines, cannabinoids, cocaine metabolites, opiates, and phencyclidine, and each assay has a specific preset cutoff concentration. Accuracy is > 99% for reporting positive or negative results for samples with 200% or 50%, r...

متن کامل

Confirmation of THC in Oral Fluids Using High-Resolution 2-D GC/MS

Oral fluids are being used as an alternative matrix to urine for drug testing. Oral fluid is considered to be less invasive and much more difficult to adulterate than urine samples. Samples of oral fluid are typically screened by an ELISA immunoassay method. Those found to be positive must be further analyzed with a mass spectrometry confirmation technique. The confirmation technique must be ab...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of analytical toxicology

دوره 22 1  شماره 

صفحات  -

تاریخ انتشار 1998