Agreement of Anterior Segment Parameters Obtained From Swept-Source Fourier-Domain and Time-Domain Anterior Segment Optical Coherence Tomography
نویسندگان
چکیده
Purpose To assess the interdevice agreement between swept-source Fourier-domain and time-domain anterior segment optical coherence tomography (AS-OCT). Methods Fifty-three eyes from 41 subjects underwent CASIA2 and Visante OCT imaging. One hundred eighty-degree axis images were measured with the built-in two-dimensional analysis software for the swept-source Fourier-domain AS-OCT (CASIA2) and a customized program for the time-domain AS-OCT (Visante OCT). In both devices, we examined the angle opening distance (AOD), trabecular iris space area (TISA), angle recess area (ARA), anterior chamber depth (ACD), anterior chamber width (ACW), and lens vault (LV). Bland-Altman plots and intraclass correlation (ICC) were performed. Orthogonal linear regression assessed any proportional bias. Results ICC showed strong correlation for LV (0.925) and ACD (0.992) and moderate agreement for ACW (0.801). ICC suggested good agreement for all angle parameters (0.771-0.878) except temporal AOD500 (0.743) and ARA750 (nasal 0.481; temporal 0.481). There was a proportional bias in nasal ARA750 (slope 2.44, 95% confidence interval [CI]: 1.95-3.18), temporal ARA750 (slope 2.57, 95% CI: 2.04-3.40), and nasal TISA500 (slope 1.30, 95% CI: 1.12-1.54). Bland-Altman plots demonstrated in all measured parameters a minimal mean difference between the two devices (-0.089 to 0.063); however, evidence of constant bias was found in nasal AOD250, nasal AOD500, nasal AOD750, nasal ARA750, temporal AOD500, temporal AOD750, temporal ARA750, and ACD. Among the parameters with constant biases, CASIA2 tends to give the larger numbers. Conclusions Both devices had generally good agreement. However, there were proportional and constant biases in most angle parameters. Thus, it is not recommended that values be used interchangeably.
منابع مشابه
Agreement between Gonioscopic Examination and Swept Source Fourier Domain Anterior Segment Optical Coherence Tomography Imaging
Purpose. To evaluate interobserver, intervisit, and interinstrument agreements for gonioscopy and Fourier domain anterior segment optical coherence tomography (FD ASOCT) for classifying open and narrow angle eyes. Methods. Eighty-six eyes with open or narrow anterior chamber angles were included. The superior angle was classified open or narrow by 2 of 5 glaucoma specialists using gonioscopy an...
متن کاملComparative Study of Anterior Eye Segment Measurements with Spectral Swept-Source and Time-Domain Optical Coherence Tomography in Eyes with Corneal Dystrophies
PURPOSE To compare anterior eye segment measurements and morphology obtained with two optical coherence tomography systems (TD OCT, SS OCT) in eyes with corneal dystrophies (CDs). METHODS Fifty healthy volunteers (50 eyes) and 54 patients (96 eyes) diagnosed with CD (epithelial basement membrane dystrophy, EBMD = 12 eyes; Thiel-Behnke CD = 6 eyes; lattice CD TGFBI type = 15 eyes; granular CD ...
متن کاملAnterior segment imaging in glaucoma: An updated review
Anterior segment imaging allows for an objective method of visualizing the anterior segment angle. Two of the most commonly used devices for anterior segment imaging include the anterior segment optical coherence tomography (AS-OCT) and the ultrasound biomicroscopy (UBM). AS-OCT technology has several types, including time-domain, swept-source, and spectral-domain-based configurations. We perfo...
متن کاملAdvances and Clinical Applications of Anterior Segment Imaging Techniques
With the rapid development of computer science and technologies in recent years, there has been dramatic advance in anterior segment imaging. As we have mentioned in the call for papers for this special issue, manuscripts are covering the topics of anterior segment imaging techniques including anterior segment optical coherence tomography (OCT), specular microscopy, corneal topography, confocal...
متن کاملUtility of anterior segment swept-source optical coherence tomography for imaging eyes with antecedent ocular trauma
Purpose To evaluate the utility of swept source optical coherence tomography (SS-OCT) for the analysis of anterior segment structures in cases with previous traumatic eye injuries. Observations We report three eyes of three patients with anterior segment traumatic eye injury and highlight the role of SS-OCT in their evaluation and management. This technology enabled us to visualize the struct...
متن کامل