Multi-echo IDEAL Cardiac Water-Fat Imaging
نویسندگان
چکیده
Introduction In cardiac magnetic resonance imaging, separation of water and fat signals is important for improved visualization of enhancing myocardium and pericardial disease, as well as direct visualization of fat in pathologies such as arrythmogenic right ventricular cardiomyopathy (ARVC), fatty tumors, and fatty infiltration of chronic myocardial infarctions. Fat can appear similar to slow blood flow with some methods, and fat may obscure enhancing myocardium, particularly epicardial enhancement that can occur in myocarditis, pericarditis, and non-ischemic cardiomyopathies. Chemical shift based water-fat separation methods have been used to provide separate water and fat images in balanced-SSFP CINE imaging [1] and myocardial viability imaging [2,3]. Here, we demonstrate the ability of IDEAL [1,4] to decompose fat and water signals in cardiac imaging with a gated multi-echo segmented gradient-echo sequence. Included are the use of near-optimal TE increments that maximize SNR performance [4] and a modified version of an advanced field-map estimation method with region-growing [5]. Methods A multi-echo gradient echo sequence with 3 echoes acquired with positive (fly-back) readouts was implemented with and without inversion recovery (IR) at 1.5T (Signa HDx, GE Healthcare, Milwaukee, WI). Parameters included: TR/TE1/ΔTE/BW = 7.2-7.8/1.51.7/1.9-2.0/±100kHz; FOV/Matrix/Slice Thickness = 35cm/192x192/6-8mm; 16-32 views per segment; and one segment in each or every second cardiac cycle, for a duration of 13-25 cardiac cycles per slice. In addition, a T2-preparation sequence (T2-Prep) with or without IR was available [6,7]. The technique was demonstrated in patients and volunteers for several clinical imaging situations. Results An example of “rule-out” suspected ARVC is shown in Figure 1, acquired in diastole without IR. A conventional double-IR blackblood sequence (1a) used for comparison showed high signal in the mid RV wall concerning for fat vs. slow flow (arrows). This was easily resolved as blood and not fat using IDEAL (1b,1c) images. Myocardial viability imaging with IR (TI=200ms) for suppression of myocardial signal is shown in Figure 2 (no infarct in this patient), with good separation of water (2a) and fat (2b) signal.
منابع مشابه
Feasibility of T2* Estimation with Chemical Shift-Based Water-Fat Separated Cardiac Imaging
Introduction T2* mapping has been investigated in cardiac MRI for imaging the myocardial BOLD effect [1] and in iron overload assessment [2]. Advanced T2* estimation techniques in the context of chemical shift-based water-fat decomposition have recently been developed for iron detection in organs such as the liver, while simultaneously estimating fat content [3]. Separately, chemical shiftbased...
متن کاملComparison of IDEAL, MultiPeak IDEAL and Fat-Saturated FSE for Imaging of Osteoarthritis (OA) Knee Patients: Initial Clinical Experience
INTRODUCTION Iterative Decomposition of Water and Fat with Echo Asymmetry and Least-Squares Estimation (IDEAL) is a promising MRI technique for robust fat and water separation (1). It was found with higher fluid/cartilage contrast-to-noise ratio efficiency and fat-suppression quality than FSE imaging in ankle. The goal of this study is to assess image quality, fat suppression and fat-water sepa...
متن کاملCardiac Imaging with Chemical Shift Based Water-Fat Separation at 3T
Introduction Chemical shift based water-fat separation techniques have been demonstrated for 1.5T cardiac imaging of several pathologies, including fatty infiltration of chronic infarction; arrhythmogenic right ventricular cardiomyopathy (ARVC); fatty tumors; and conditions such as myocarditis and pericarditis [1-4]. Higher field strengths (3T) are now often used in cardiac MRI due to improved ...
متن کاملIDEAL at 7T in Mice Using Asymmetric Spin Echo and Gradient Echo Acquisitions
An asymmetric spin echo (aSE) technique was developed to produce uniform, robust fat-water separation in mice at 7T using Iterative Decomposition of Water and Fat with Echo Asymmetry and Least-squares estimation method (IDEAL). The aSE technique had superior image quality as compared to gradient echo IDEAL estimation. Both the spin echo and gradient echo IDEAL techniques resulted in more accura...
متن کاملIterative decomposition of water and fat with echo asymmetry and least-squares estimation (IDEAL) fast spin-echo imaging of the ankle: initial clinical experience.
OBJECTIVE Reliable, uniform fat suppression is important. Multiple approaches currently exist, many of which suffer from either suboptimal signal-to-noise ratio (SNR), or the inability to obtain consistent fat suppression around the ankle joint. Our purpose was to test iterative decomposition of water and fat with echo asymmetry and the least-squares estimation (IDEAL) method in combination wit...