Improving Atomic Force Microscopy Imaging by a Direct Inverse Asymmetric PI Hysteresis Model
نویسندگان
چکیده
A modified Prandtl-Ishlinskii (PI) model, referred to as a direct inverse asymmetric PI (DIAPI) model in this paper, was implemented to reduce the displacement error between a predicted model and the actual trajectory of a piezoelectric actuator which is commonly found in AFM systems. Due to the nonlinearity of the piezoelectric actuator, the standard symmetric PI model cannot precisely describe the asymmetric motion of the actuator. In order to improve the accuracy of AFM scans, two series of slope parameters were introduced in the PI model to describe both the voltage-increase-loop (trace) and voltage-decrease-loop (retrace). A feedforward controller based on the DIAPI model was implemented to compensate hysteresis. Performance of the DIAPI model and the feedforward controller were validated by scanning micro-lenses and standard silicon grating using a custom-built AFM.
منابع مشابه
Model Development and Inverse Compensator Design for High Speed Nanopositioning
This paper focuses on the development of constitutive models, commensurate system models, and inverse compensator construction for high speed nanopositioning in atomic force microscopes (AFM). All current AFM employ either stacked or cylindrical piezoceramic actuators for both longitudinal and transverse positioning of the sample. An inherent property of these materials is the presence of hyste...
متن کاملIdentification, Control and Hysteresis Compensation of a 3 Dof Metrological Afm
Atomic Force Microscopes (AFMs) are widely used for the investigation of samples at the nanometer scale. The metrological AFM used in this work uses a 3 degrees-of-freedom (DOFs) stage, driven by piezo-stack actuators, for sample manipulation in combination with a fixed cantilever. The piezostack actuators suffer from hysteresis, which acts as a nonlinear disturbance on the system and/or can ch...
متن کاملA Preisach Model for Quantifying Hysteresis in an Atomic Force Microscope
Atomic force microscopes employ stacked or cylindrical piezoceramic actuators to achieve sub-angstrom resolution. While these devices produce excellent set-point accuracy, they exhibit hysteresis and constitutive nonlinearities even at low drive levels. Feedback mechanisms can mitigate the deleterious effects of these nonlinearities for low frequency operation but such techniques fail at higher...
متن کاملSensitivity Analysis of Frequency Response of Atomic Force Microscopy in Liquid Environment on Cantilever's Geometrical Parameters
In this paper, the non-linear dynamic response of rectangular atomic force microscopy in tapping mode is considered. The effect of cantilever’s geometrical parameters (e.g., cantilever length, width, thickness, tip length and the angle between the cantilever and the sample's surface in liquid environment has been studied by taking into account the interaction forces. Results indicate that the r...
متن کاملAdvanced Control of Atomic Force Microscope for Faster Image Scanning
In atomic force microscopy (AFM), the dynamics and nonlinearities of its nanopositioning stage are major sources of image distortion, especially when imaging at high scanning speed. This chapter discusses the design and experimental implementation of an observer-based model predictive control (OMPC) scheme which aims to compensate for the effects of creep, hysteresis, cross-coupling, and vibrat...