Development and validation of a new automatic algorithm for quantification of left ventricular volumes and function in gated myocardial perfusion SPECT using cardiac magnetic resonance as reference standard.
نویسندگان
چکیده
BACKGROUND By gating image acquisition in myocardial perfusion SPECT (MPS) to ECG, left ventricular (LV) volumes and function can be determined. Several previous studies have shown that existing MPS software packages underestimate LV volumes compared to cardiac magnetic resonance (CMR). The aim of this study was therefore to develop a new LV segmentation algorithm for gated MPS using CMR as reference standard. METHODS AND RESULTS A total of 126 patients with suspected coronary artery disease, who underwent both gated MPS and CMR were retrospectively included. The proposed LV segmentation algorithm (Segment) was trained in 26 patients, and tested in 100 patients in comparison to four commercially available MPS software packages (QGS, MyoMetrix, ECTb, and Exini) using CMR as reference standard. Mean bias ± SD between MPS and CMR was for EDV -5% ± 12%, -43% ± 8%, -40% ± 8%, -42% ± 9%, -32% ± 7%, for ESV 0% ± 17%, -41% ± 16%, -34% ± 15%, -54% ± 13%, -41% ± 10%, for EF -2% ± 13%, -1% ± 14%, -7% ± 15%, 17% ± 16%, 10% ± 17% for Segment, QGS, MyoMetrix, ECTb, and Exini, respectively, and for LVM 3% ± 18%, 33% ± 25%, 37% ± 24% for Segment, QGS, and ECTb, respectively. Correlation between MPS by Segment and CMR were for EDV R (2) = 0.89, for ESV R (2) = 0.92, for EF R (2) = 0.69, and for LVM R (2) = 0.72, with no difference compared to the correlation between the other MPS software packages and CMR (EDV R (2) = 0.86-0.92, ESV R (2) = 0.91-0.93, EF R (2) = 0.64-0.65, and LVM R (2) = 0.68-0.70). CONCLUSION The Segment software quantifies LV volumes and EF by MPS with similar correlation and a low bias compared to other MPS software packages, using CMR as reference standard. Hence, the Segment software shows potential to provide clinically relevant volumes and functional values from MPS.
منابع مشابه
Development and validation of a new automatic algorithm for quantification of left ventricular volumes and function in gated myocardial perfusion SPECT using cardiac magnetic resonance as reference
Citation for published version (APA): Fransson, H., Hedeer, F., Arévalo, C., Carlsson, M., Engblom, H., Ubachs, J., ... Heiberg, E. (2011). Development and validation of a new automatic algorithm for quantification of left ventricular volumes and function in gated myocardial perfusion SPECT using cardiac magnetic resonance as reference standard. Journal of Nuclear Cardiology, 18(5), 874-885. DO...
متن کاملThe prognostic value of stress/rest gated myocardial perfusion SPECT in patients with known or suspected coronary artery disease
Introduction:Gated myocardial perfusion SPECT has diagnostic and prognostic values in coronary artery disease (CAD). We tried to determine prognostic values of the left ventricular perfusion & functional indices as well as transient left ventricular dilation (TID) derived from gated myocardial perfusion SPECT. Methods: 1820patients who underwent gated myo...
متن کاملDiastolic left ventricular functional indices: Comparison between gated myocardial perfusion SPECT and echocardiography
Introduction: Many patients have only diastolic dysfunction without any abnormality in systolic function. We compared cardiac diastolic parameters obtained by gated myocardial perfusion SPECT (Gated SPECT) with those evaluated by echocardiography. Methods: Forty-nine patients (aged 37-85 years, 19 males and 30 women) underwent Gated SPECT and echocardiogr...
متن کاملDipyridamole stress and rest gated 99mTc-sestamibi myocardial perfusion SPECT: left ventricular function indices and myocardial perfusion findings
Introduction: We investigated the difference in left ventricular ejection fraction (LVEF) and end-systolic volume(ESV) measured by gated myocardial perfusion SPECT (GSPECT) in the post-dipyridamole stress and rest periods, and compared the results with the perfusion patterns found in the conventional non-gated tomograms. Methods: 297 consecutive patients were studie...
متن کاملNormal values of left ventricular functional indices in gated 99mTc-MIBI myocardial perfusion SPECT
Introduction: Although left ventricular(LV) function parameters measured by gated myocardial perfusion SPECT (GSPECT) have been validated, experimental data have revealed that the calculated the LV function parameters using GSPECT are affected by patient populations as well as particular acquisition and processing conditions. We tried to determine the normal values ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of nuclear cardiology : official publication of the American Society of Nuclear Cardiology
دوره 18 5 شماره
صفحات -
تاریخ انتشار 2011