An inexpensive dual-chamber particle monitor: laboratory characterization.
نویسندگان
چکیده
In developing countries, high levels of particle pollution from the use of coal and biomass fuels for household cooking and heating are a major cause of ill health and premature mortality. The cost and complexity of existing monitoring equipment, combined with the need to sample many locations, make routine quantification of household particle pollution levels difficult. Recent advances in technology, however, have enabled the development of a small, portable, data-logging particle monitor modified from commercial smoke alarm technology that can meet the needs of surveys in the developing world at reasonable cost. Laboratory comparisons of a prototype particle monitor developed at the University of California at Berkeley (UCB) with gravimetric filters, a tapered element oscillating microbalance, and a TSI DustTrak to quantify the UCB particle monitor response as a function of both concentration and particle size and to examine sensor response in relation to changes in temperature, relative humidity, and elevation are presented here. UCB particle monitors showed good linearity in response to different concentrations of laboratory-generated oleic acid aerosols with a coarse (mass median diameter, 2.1 microm) and fine (mass median diameter, 0.27-0.42 microm) size distributions (average r2 = 0.997 +/- 0.005). The photoelectric and ionization chamber showed a wide range of responses based on particle size and, thus, require calibration with the aerosol of interest. The ionization chamber was five times more sensitive to fine rather than coarse particles, whereas the photoelectric chamber was five times more sensitive to coarse than fine. The ratio of the response between the two sensors has the potential for mass calibration of individual data points based on estimated parameters of the size distribution. The results demonstrate the significant potential of this monitor, which will facilitate the evaluation of interventions (improved fuels, stoves, and ventilation) on indoor air pollution levels and research on the impacts of indoor particle levels on health in developing countries.
منابع مشابه
An inexpensive light-scattering particle monitor: field validation.
We have developed a small, light, passive, inexpensive, datalogging particle monitor called the "UCB" (University of California Berkeley particle monitor). Following previously published laboratory assessments, we present here results of tests of its performance in field settings at high particle concentrations. We demonstrate the mass sensitivity of the UCB in relation to gravimetric filter-ba...
متن کاملResponse Characterization of an Inexpensive Aerosol Sensor
Inexpensive aerosol sensors have been considered as a complementary option to address the issue of expensive but low spatial coverage air quality monitoring networks. However, the accuracy and response characteristics of these sensors is poorly documented. In this study, inexpensive Shinyei PPD42NS and PPD60PV sensors were evaluated using a novel laboratory evaluation method. A continuously cha...
متن کاملAn inexpensive particle monitor for smoker behaviour modification in homes.
OBJECTIVE To compare the response of a new particle counting instrument (Dylos DC1700) with that produced by a device (TSI Sidepak Personal Aerosol Monitor AM510) commonly used to measure PM2.5 in settings where secondhand smoke (SHS) is encountered. METHODS Controlled chamber experiments with different SHS concentrations were generated by burning a cigarette for varying time periods and runn...
متن کاملDesign of an Inhalation Chamber to Expose Laboratory Animals with Suspended Particulate Matter by Numerical Simulation Method
Introduction: Usually, in the toxicological studies of airborne particulate pollutants, inhalation exposure chambers are used for providing and distributing the test atmosphere uniformly and stability in the respiratory zone of laboratory animals. The purpose of this study was to design, evaluate and optimize a whole-body exposure chamber, specifically for small laboratory animals exposed to pa...
متن کاملPreliminary Characterization Tests of Detectors of on-Line Monitor Systems of the Italian National Center of Oncological Hadron-Therapy (CNAO)
Introduction Hadron-therapy is an effective technique used to treat tumors that are located between or nearby vital organs. The Italian National Center of Oncological Hadron-therapy (CNAO) has been realized as the first facility in Italy to treat very difficult tumors with protons and Carbon ions. The on-line monitor system for CNAO has been developed by the Department of Physics of the Univers...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Journal of the Air & Waste Management Association
دوره 56 6 شماره
صفحات -
تاریخ انتشار 2006