Reoperation for mid-term failure of a freestyle bioprosthesis using a full root technique --a surgical case--.
نویسندگان
چکیده
The aortic root of a 30-year-old man was replaced with a Freestyle stentless aortic bioprosthesis for aortic regurgitation associated with annuloaortic ectasia. His clinical course was uneventful, and he was discharged without complications. Three years and six months after surgery, he presented with a high fever. Four years after surgery, transthoracic echocardiography revealed severe aortic regurgitation. We performed exploratory surgery and discovered a torn left coronary cusp of the Freestyle bioprosthesis. Organized vegetation was adherent to the left coronary cusp leaflet. The non-coronary cusp and the right coronary cusp were normal. The diagnosis was aortic regurgitation due to valve failure related to infective endocarditis. Consequently, we reconstructed the aortic root with a composite graft (26-mm Valsalva graft and a 21-mm ON-X mechanical valve).
منابع مشابه
Long-term clinical outcomes 15 years after aortic valve replacement with the Freestyle stentless aortic bioprosthesis.
BACKGROUND The Freestyle stentless aortic root bioprosthesis has excellent hemodynamics and durability through 10 years. The purpose of this report is to present clinical outcomes in a large multicenter cohort through 15 years. METHODS The multicenter evaluation of the Freestyle valve began in 1992 at 21 centers in North America and Europe. In 1997, a long-term study continued, including 725 ...
متن کاملStructural deterioration of the Freestyle aortic valve: mode of presentation and mechanisms.
BACKGROUND Structural valve deterioration is the major cause of bioprosthetic valve failure. Because of the unique design features and anti-calcification treatment of the Freestyle (Medtronic Inc, Minneapolis, Minn) stentless bioprosthesis, development of structural valve deterioration may differ in comparison with other bioprosthetic valves. This study evaluates the mechanisms and clinical pre...
متن کاملBioprosthetic valve durability after stentless aortic valve replacement: the effect of implantation technique.
BACKGROUND The Freestyle stentless bioprosthesis (FSB) (Medtronic Inc, Minneapolis, MN) is implanted using 2 techniques-subcoronary or aortic root replacement. Our objective was to determine whether the implantation technique had an impact on late reoperation for structural valve deterioration (SVD). METHODS Between 1993 and 2013, 531 patients underwent aortic valve replacement (AVR) or aorti...
متن کاملLong-term clinical and echocardiographic follow-up of the Freestyle stentless aortic bioprosthesis.
BACKGROUND Stentless aortic bioprostheses were designed to provide enhanced hemodynamic performance and potentially greater longevity. The present report describes the outcomes of patients with the Freestyle stentless bioprosthesis followed for ≤18 years. METHODS AND RESULTS Between 1993 and 2011, 430 patients underwent primary aortic valve replacement with a Freestyle bioprosthesis in the su...
متن کاملAortic Root Replacement for Pseudoaneurysm Arising from Freestyle Aortic Bioprosthesis.
A 70-year-old female presented with a new systolic murmur and shortness of breath three years after undergoing aortic root replacement using a 27-mm Medtronic Freestyle® stentless full root bioprosthesis (Medtronic, Inc., Fridley, Minnesota, USA). A large complex aortic root pseudoaneurysm was identified on contrasted computed tomography of the chest and transesophageal echocardiogram. We descr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia
دوره 17 3 شماره
صفحات -
تاریخ انتشار 2011