OBJECTIVES This retrospective study was performed to provide data on ventricular tachycardias (VT) with a cycle length longer than the initially programmed tachycardia detection interval (TDI) in patients with implantable cardioverter defibrillators (ICDs). BACKGROUND It has been clinical practice to program a safety margin of 30 to 60 ms between the slowest spontaneous or inducible VT and the TDI. METHODS Baseline characteristics of 659 consecutive patients with ICDs were prospectively; follow-up information was retrospectively collected. RESULTS During a mean follow-up of 31+/-23 months, 377 patients (57.2%) had at least one recurrent VT or ventricular fibrillation; 47 patients (7.1%) suffered 61 VTs above the TDI. The risk of a VT above the TDI ranged between 2.7% and 3.5% per year during the first four years after ICD implantation. The difference between the cycle length of the slowest VT before ICD implantation, spontaneous or induced, and the first VT above TDI was 108+/-58 ms. Fifty-four VTs (88.5%) above the TDI were associated with significant clinical symptoms (angina or palpitation 63.9%, heart failure 6.6% and syncope 8.2%). Six patients (9.8%) had to be resuscitated. Kaplan-Meyer analysis identified New York Heart Association class II or III (p = 0.021), ejection fraction < 0.40 (p = 0.027), spontaneous (p<0.001) or inducible (p<0.001) monomorphic VTs and the use of class III antiarrhythmic drugs (amiodarone, p<0.001; sotalol, p = 0.004) as risk predictors of VTs above the TDI. The risk of recurrent VTs above TDI was 11.8%, 12.5% and 26.6% during the first, second and third year after first VT above TDI, respectively. CONCLUSIONS The risk of VTs above the TDI is significantly increased in some patients, and many VTs above TDI cause significant clinical symptoms. A larger safety margin between spontaneous or inducible VTs and the TDI seems to be necessary in selected patients. This is in conflict with an increased risk of inadequate episodes and demands highly specific and sensitive detection algorithms in these patients.
منابع مشابه
Ventriculoatrial conduction metrics for classification of ventricular tachycardia with 1:1 retrograde conduction in dual-chamber sensing implantable cardioverter defibrillators.
The introduction of dual-chamber sensing in implantable cardioverter defibrillators (ICDs) has greatly reduced the incidence of false detection due to supraventricular tachycardias. The remaining arrhythmias which serve to confound classification are supraventricular tachycardias (SVTs) with 1:1 anterograde conduction and ventricular tachycardias (VT) with 1:1 retrograde conduction. An algorith...
متن کاملProarrhythmic effect of pacemaker stimulation in patients with implanted cardioverter-defibrillators.
BACKGROUND We sought to determine the potential of right ventricular VVI backup pacing to induce ventricular tachyarrhythmias in patients with implanted cardioverter-defibrillators. METHODS AND RESULTS All consecutive patients presenting exclusively with pacemaker-induced tachycardias (PITs) were included in a prospective study using a crossover protocol. Patients were randomized to either gr...
متن کاملThe 1+1 trial: a prospective trial of a dual- versus a single-chamber implantable defibrillator in patients with slow ventricular tachycardias.
BACKGROUND The tachycardia detection interval (TDI) in implantable cardioverter/defibrillators (ICDs) is conventionally programmed according to the slowest documented ventricular tachycardia (VT), with a safety margin of 30 to 60 ms. With this margin, VTs above the TDI may occur. However, longer TDIs are associated with an increased risk of inappropriate therapy. We hypothesized that patients w...
متن کاملClinical relevance of slow ventricular tachycardia in heart failure patients with primary prophylactic implantable cardioverter defibrillator indication.
AIMS Implantable cardioverter defibrillators (ICDs) have shown to reduce all-cause mortality in heart failure patients. In SCD-HeFT study, ICDs were programmed with a detection zone of ≥ 187 b.p.m. Thus, the incidence and clinical significance of slower ventricular tachycardias (VTs) in these patients remains largely unknown, though clinically important for device selection, programming, and fo...
متن کاملIrregular ventricular tachycardia underdetected by implantable cardioverter defibrillator device.
A case of sustained monomorphic ventricular tachycardia underdetected by a single chamber implantable cardioverter defibrillator because of RR interval irregularity is presented. The programmed stability criterion is responsible for the underdetection. Special attention must be paid when it comes to programming this detection parameter.
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of the American College of Cardiology
دوره 36 2 شماره
صفحات -
تاریخ انتشار 2000