Cardiac output monitoring by pressure recording analytical method in cardiac surgery.

نویسندگان

  • Pierpaolo Giomarelli
  • Bonizella Biagioli
  • Sabino Scolletta
چکیده

OBJECTIVE A less-invasive method has been developed that may provide an alternative to monitor cardiac output from arterial pressure: beat-to-beat values of cardiac output can be obtained by pressure recording analytical method (PRAM). The purpose of this study was to assess the reliability of cardiac output determination by PRAM in cardiac surgery. METHODS Cardiac output was measured in 28 patients undergoing coronary artery bypass grafting at 15 min after anaesthesia induction, 30 min after extracorporeal circulation, 1 and 3 h after arrival in the intensive care unit using thermodilution (ThD) method through a pulmonary artery catheter and PRAM. ThD cardiac output was calculated as the mean of five separate measurements. PRAM provided beat-by-beat cardiac output data continuously throughout the study and the cardiac output values displayed on a dedicated personal computer at each time point were recorded. Correlations were calculated and differences were compared by Bland-Altman analysis. RESULTS A total of 112 measurements were obtained. Cardiac output ranged from 2.3 to 7.4 l/min, and a good linear correlation (R2=0.78, P<0.0001) was found between ThD and PRAM. The highest degree of correlation (R2=0.86) was obtained at 3 h after arrival in the intensive care unit. The lower degree of correlation (R2=0.70) was obtained 30 min after extracorporeal circulation. At Bland-Altman analysis, the overall estimates of cardiac output measured by PRAM closely agreed with ThD (mean difference, 0.027; standard deviation, 0.43; limits of agreement, -0.83 and +0.89). CONCLUSIONS Under the studied conditions, our results demonstrate good agreement between PRAM data and ThD measurements, and this new method has shown to be accurate for real-time monitoring of cardiac output in cardiac surgery. Further studies will be required to assess this method in higher-risk patients and in the setting of haemodynamic instability or arrhythmias. This is the first study designed to assess the accuracy of PRAM in cardiac surgery.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Lack of agreement between pulmonary arterial thermodilution cardiac output and the pressure recording analytical method in postoperative cardiac surgery patients.

BACKGROUND Pulse-contour analysis method (PCM) cardiac output (CO) monitors are increasingly used for CO monitoring during anaesthesia and in the critically ill. Very recently, several systems have been introduced that do not need calibration; among them the pressure recording analytical method (PRAM). Sparse data comparing the accuracy of the PRAM-CO with conventional thermodilution CO (ThD-CO...

متن کامل

A System for Continuous Estimating and Monitoring Cardiac Output via Arterial Waveform Analysis

Background: Cardiac output (CO) is the total volume of blood pumped by the heart per minute and is a function of heart rate and stroke volume. CO is one of the most important parameters for monitoring cardiac function, estimating global oxygen delivery and understanding the causes of high blood pressure. Hence, measuring CO has always been a matter of interest to researchers and clinicians. Sev...

متن کامل

Most Care&reg;: a minimally invasive system for hemodynamic monitoring powered by the Pressure Recording Analytical Method (PRAM)

Invasive hemodynamic monitoring is a cornerstone of the care of critically ill and hemodynamically unstable patients in both intensive care units and operating rooms. The assessment of cardiac output by means of the pulmonary artery catheter has been considered the clinical gold standard. Nevertheless, several concerns have been raised regarding its invasiveness, usefulness, and associated comp...

متن کامل

Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states.

BACKGROUND Cardiac output (CO) can be measured using the pressure recording analytical method (PRAM), which is a new, less invasive technique allowing beat-by-beat stroke volume monitoring from the pressure signals recorded in femoral or radial arteries. METHODS We investigated PRAM by comparing its cardiac output (PRAM-CO) with paired measurements obtained by electromagnetic flowmetry (EM-CO...

متن کامل

Comparison between pressure-recording analytical method (PRAM) and femoral arterial thermodilution method (FATD) cardiac output monitoring in an infant animal model of cardiac arrest

BACKGROUND The pressure-recording analytical method is a new semi-invasive method for cardiac output measurement (PRAM). There are no studies comparing this technique with femoral artery thermodilution (FATD) in an infant animal model. METHODS A prospective study was performed using 25 immature Maryland pigs weighing 9.5 kg. Fifty-eight simultaneous measurements of cardiac index (CI) were mad...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery

دوره 26 3  شماره 

صفحات  -

تاریخ انتشار 2004