Instrument intercomparison in the high-energy mixed field at the CERN-EU reference field (CERF) facility.
نویسندگان
چکیده
This paper discusses an intercomparison campaign performed in the mixed radiation field at the CERN-EU (CERF) reference field facility. Various instruments were employed: conventional and extended-range rem counters including a novel instrument called LUPIN, a bubble detector using an active counting system (ABC 1260) and two tissue-equivalent proportional counters (TEPCs). The results show that the extended range instruments agree well within their uncertainties and within 1σ with the H*(10) FLUKA value. The conventional rem counters are in good agreement within their uncertainties and underestimate H*(10) as measured by the extended range instruments and as predicted by FLUKA. The TEPCs slightly overestimate the FLUKA value but they are anyhow consistent with it when taking the comparatively large total uncertainties into account, and indicate that the non-neutron part of the stray field accounts for ∼30 % of the total H*(10).
منابع مشابه
The CERN-EU high-energy reference field (CERF) facility for dosimetry at commercial flight altitudes and in space.
A reference facility for the calibration and intercomparison of active and passive detectors in broad neutron fields has been available at CERN since 1992. A positively charged hadron beam (a mixture of protons and pions) with momentum of 120 GeV/c hits a copper target, 50 cm thick and 7 cm in diameter. The secondary particles produced in the interaction traverse a shield, at 90 degrees with re...
متن کاملOrganisation Europeenne Pour La Recherche Nucleaire European Organization for Nuclear Research
A comparison of two types of dose equivalent meters, a recombination chamber REM-2 and a tissue-equivalent proportional counter HANDI, were carried out at the CERN-EC high energy reference field facility, CERF, in October 2001. Here, the mixed high-energy radiation field in commercial flight altitudes can be simulated with dose equivalent rates up to 2 orders of magnitude higher. The total abso...
متن کاملA TLD-based personal dosemeter system for aircrew monitoring.
Evaluation of LiF:Mg,Ti thermoluminescence dosemeters (TLDs) according to the high-temperature ratio (HTR) method enables the determination of the dose-average linear energy transfer (LET), the mean quality factor and the dose equivalent in mixed radiation fields of unknown composition. The neutron contribution is assessed by the Extended Pair method calibrated in the CERN-EU High-Energy Refere...
متن کاملPerformance Requirements for Monitoring Pulsed, Mixed Radiation Fields around High-energy Accelerators
RADIATION FIELDS AND MONITORS In preparation of the installation of a radiation monitoring system for the future LHC and its injectors comprehensive studies were performed to evaluate the suitability of different existing monitors for radiation protection measurements in pulsed and complex, mixed radiation fields. Different ionisation chambers were exposed to a mixed, high-energy reference fiel...
متن کاملBenchmark studies of induced radioactivity produced in LHC materials, Part II: Remanent dose rates.
A new method to estimate remanent dose rates, to be used with the Monte Carlo code FLUKA, was benchmarked against measurements from an experiment that was performed at the CERN-EU high-energy reference field facility. An extensive collection of samples of different materials were placed downstream of, and laterally to, a copper target, intercepting a positively charged mixed hadron beam with a ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Radiation protection dosimetry
دوره 161 1-4 شماره
صفحات -
تاریخ انتشار 2014