A case report of progressive obstruction of Ex-PRESS miniature glaucoma shunt after transient flat anterior chamber and treatment using Nd:YAG laser

نویسندگان

  • Masaki Tanito
  • Ichiya Sano
  • Akihiro Ohira
چکیده

BACKGROUND We report a case of Ex-PRESS miniature glaucoma shunt obstruction resulting from progressive iris synechial formation after transient anterior chamber shallowing. CASE PRESENTATION A 68-year-old woman with pseudoexfoliation glaucoma in her right eye underwent filtration surgery with implantation of the Ex-PPESS shunt (model P-50, Alcon Japan, Tokyo, Japan) in combination with intra-surgical 0.04% mitomycin C use. After the anterior chamber injection of viscoelastic material and 100% sulfur hexafluoride gas for treatment of early postoperative over filtration, the intraocular pressure (IOP) was controlled between 9 and 12 mmHg. On postoperative day 121, gonioscopy showed that synechial formation around the shunt obstructed the axial port leaving the relief port opened. On postoperative day 274, the intraocular pressure increased to 40 mmHg and synechiae obstructed both the axial and relief ports. Dispersion of iris tissue by neodymium:yttrium-aluminium-garnet (Nd:YAG) laser (2 mJ, one shot to each port) opened both ports and immediately lowered the IOP, leaving peripheral anterior synechiae around the shunt. Up to postoperative day 400, the IOP was controlled between 13 and 15 mmHg, and the cystic bleb was maintained. CONCLUSION The synechiae formed gradually extends around the shunt's shaft and can result in later external obstruction of the relief port. The current case requires further follow-up since synechiae remaining around the shaft can cause future obstruction. We emphasize the fact that, if the iris synechiae to the shunt once formed, it can progress and obstruct the shunt ports later.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Fractured Anterior Chamber Intraocular Lens (ACIOL) Complicating Nd: YAG Laser for Peripheral Iridotomy

Laser peripheral iridotomy is the procedure of choice for the treatment of angle-closure glaucoma caused by relative or absolute pupillary block. Nd: YAG laser iridotomy has been reported to have several complications such as Iris bleeding, hyphema, transient IOP elevation, intraocular inflammation, choroidal, retinal detachment and vitreous hemorrhage. We report a case of a 74 year old lady on...

متن کامل

New ab interno technique for removal of iris-embedded EX-PRESS shunt and chronic eye pain caused by shunt malpositioning.

The EX-PRESS shunt (Alcon Laboratory) is a stainless steel implant designed to shunt aqueous from the anterior chamber to the subconjunctival space.Thedevice isplacedundera scleral flap,with the lumenideallypositionedparallel to the iris. A properly positioned EX-PRESS shunt creates acontrolledcommunicationbetween the anterior chamber and the bleb.1 Several shunt-specific complications havebeen...

متن کامل

Ex-PRESS miniature glaucoma shunt in treatment of refractory glaucoma.

Refractory glaucoma in a complicated type of glaucoma of different ethyologies with one same characteristic--intraocular pressure of great resistance to therapy. There are different methods of treatment in such glaucomas, primary surgical options. Ex-PRESS miniature glaucoma shunt implantation was our treatment of choice. In our group of patients we achieved stabile intraocular pressure values ...

متن کامل

Impending extrusion of Ex-PRESS shunt treated by shunt-position adjustment: a case report

BACKGROUND To report a case of impending extrusion of Ex-PRESS shunt treated by shunt-position adjustment. CASE PRESENTATION A 56-year-old Asian woman presented with impending extrusion and malposition of Ex-PRESS shunt in her left eye. The bleb of the left eye was shallow and diffuse. In the past, the patient was treated by Ex-PRESS shunt implantation under the scleral flap in both eyes. The...

متن کامل

Secondary glaucoma due to anterior chamber lipoma and uveitis in a Pekingese dog

The present study introduces a rare case with secondary glaucoma due to anterior chamber lipoma and uveitis in a Pekingese dog in Iran. A four-year-old castrated male Pekingese dog was referred to the private clinic with impaired vision of the left eye, pain, rubbing the left eye onto carpet, presence of a milkycoloured mass into the anterior chamber, and loss of appetite. The clinical examinat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2015