Airway humidification during high-frequency percussive ventilation.

نویسندگان

  • Patrick F Allan
  • Michael J Hollingsworth
  • Gordon C Maniere
  • Anthony K Rakofsky
  • Kevin K Chung
  • Gregory A Naworol
  • John A Ward
  • Michelle Perello
  • Michael J Morris
چکیده

BACKGROUND We were concerned about the risk of inadequate humidification during high-frequency percussive ventilation (HFPV). METHODS We studied 5 humidifiers during HFPV with a lung model, at bias gas flows of 10 L/min, 30 L/min, and 50 L/min, and compared the results to those from a comparator ventilator/humidifier setup and to the minimum temperature (30 degrees C) and humidity (30 mg/L) [corrected] recommended by the American Association for Respiratory Care, at both regular room temperature and a high ambient temperature. Temperature was measured at the humidifier outflow point and at the artificial carina. Humidity was measured at the artificial carina. RESULTS Of the 7 HFPV/humidifier combinations, 2 (the MR850 at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided a carinal temperature equivalent to the comparator setup at room temperature, whereas one HFPV/humidifier combination (the ConchaTherm Hi-Flow with modified programming, at bias flows of 30 L/min and 50 L/min) provided a higher carinal temperature. At high ambient temperature, all of the setups delivered lower carinal temperature than the comparator setup. Only 2 setups (the ConchaTherm with modified programming at a bias flow of 50 L/min, and the ConchaTherm Hi-Flow with VDR nebulizer) provided carinal humidification equivalent to the comparator setup, without regard to ambient temperature; the other humidifiers were less effective. The ConchaTherm with modified programming, and the ConchaTherm with the VDR nebulizer provided the most consistent humidification. CONCLUSION HFPV's distinctive gas-flow mechanism may impair gas heating and humidification, so all humidification systems should be tested with HFPV prior to clinical use.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A new prototype of an electronic jet-ventilator and its humidification system

BACKGROUND: Adequate humidification in long-term jet ventilation is a critical aspect in terms of clinical safety. AIM: To assess a prototype of an electronic jet-ventilator and its humidification system. METHODS: Forty patients with respiratory insufficiency were randomly allocated to one of four groups. The criterion for inclusion in this study was respiratory insufficiency exhibiting a Murra...

متن کامل

The Usage of Intrapulmonary Percussive Ventilation

Mechanism of action Intrapulmonary percussive ventilation (IPV) is a form of high frequency oscillatory ventilation that can be used via mouthpiece, mask interface, or inline with artificial airway or mechanical ventilation. IPV was designed to improve mucus clearance and recruit obstructed bronchi and alveoli using a lung protective strategy; it avoids barotrauma caused by over-inflation of pr...

متن کامل

Secretion management in the mechanically ventilated patient.

Secretion management in the mechanically ventilated patient includes routine methods for maintaining mucociliary function, as well as techniques for secretion removal. Humidification, mobilization of the patient, and airway suctioning are all routine procedures for managing secretions in the ventilated patient. Early ambulation of the post-surgical patient and routine turning of the ventilated ...

متن کامل

Haemodynamics and oxygenation improvement induced by high frequency percussive ventilation in a patient with hypoxia following cardiac surgery: a case report

INTRODUCTION High frequency percussive ventilation is a ventilatory technique that delivers small bursts of high flow respiratory gas into the lungs at high rates. It is classified as a pneumatically powered, pressure-regulated, time-cycled, high-frequency flow interrupter modality of ventilation. High frequency percussive ventilation improves the arterial partial pressure of oxygen with the sa...

متن کامل

Measurement of pulsatile tidal volume, pressure amplitude, and gas flow during high-frequency percussive ventilation, with and without partial cuff deflation.

OBJECTIVE With a high-frequency percussive ventilator and a mechanical lung model, to measure tidal volume (V(T)), pulsatile pressure amplitude (difference between peak and nadir pulsatile pressure [DeltaP];), and mean airway pressure (P (aw)) at various pulsatile frequencies, pulsatile inspiratory-expiratory ratios (I:E(p)), and pressures (measured at the interface between the pulse-generator ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Respiratory care

دوره 54 3  شماره 

صفحات  -

تاریخ انتشار 2009