Accuracy of image-guided surgical systems at the lateral skull base as clinically assessed using bone-anchored hearing aid posts as surgical targets.
نویسندگان
چکیده
OBJECTIVE Image-guided surgical (IGS) technology has been clinically available for more than a decade. To date, no acceptable standard exists for reporting the accuracy of IGS systems, especially for lateral skull base applications. We present a validation method that uses the post from bone-anchored hearing aid (BAHA) patients as a target. We then compare the accuracy of 2 IGS systems-one using laser skin-surface scanning (LSSS) and another using a noninvasive fiducial frame (FF) attached to patient via dental bite-block (DBB) for registration. STUDY DESIGN Prospective. SETTING Tertiary referral center. PATIENTS Six BAHA patients who had adequate dentition for creation of a DBB. INTERVENTION(S) For each patient, a dental impression was obtained, and a customized DBB was made to hold an FF. Temporal bone computed tomographic (CT) scans were obtained with the patient wearing the DBB, FF, and a customized marker on the BAHA post. In a mock operating room, CT scans were registered to operative anatomy using 2 methods: 1) LSSS and 2) FF. MAIN OUTCOME MEASURE(S) For each patient and each system, the position of the BAHA marker in the CT scan and in the mock operating room (using optical measurement technology) was compared, and the distances between them are reported to demonstrate accuracy. RESULTS Accuracy (mean +/- standard deviation) was 1.54 +/- 0.63 mm for DBB registration and 3.21 +/- 1.02 mm for LSSS registration. CONCLUSION An IGS system using FF registration is more accurate than one using LSSS (p = 0.03, 2-sided Student's t test). BAHA patients provide a unique patient population upon which IGS systems may be validated.
منابع مشابه
The accuracy of an electromagnetic navigation system in lateral skull base approaches.
OBJECTIVES/HYPOTHESIS Image-guided optical tracking systems are being used with increased frequency in lateral skull base surgery. Recently, electromagnetic tracking systems have become available for use in this region. However, the clinical accuracy of the electromagnetic tracking system has not been examined in lateral skull base surgery. This study evaluates the accuracy of electromagnetic n...
متن کاملValidation of a Fiducial Frame System for Image-guided Otologic Surgery Utilizing Baha® Bone Screws
Image-guided otologic surgeries performed on healthy patients require non-invasive fiducial systems that achieve submillimetric target registration error (TRE). We have described such a system that attaches repeatably to a subject via a dental bite block [1]. We now describe a new method for the clinical validation of such a system that can be conducted in the laboratory with no intraoperative ...
متن کاملSubmillimetric target-registration error using a novel, non-invasive fiducial system for image-guided otologic surgery.
OBJECTIVE Otologic surgery is undertaken to treat ailments of the ear, including persistent infections, hearing loss, vertigo, and cancer. Typically performed on otherwise-healthy patients in outpatient facilities, the application of image-guided surgery (IGS) has been limited because accurate (<1 mm), non-invasive fiducial systems for otologic surgery have not been available. We now present su...
متن کاملClinical Determination of Target Registration Error of an Image- Guided Otologic Surgical System using Patients with Bone-Anchored Hearing Aids
Image guidance in otologic surgery has been thwarted by the need for a non-invasive fiducial system with target registration error (TRE) at the inner ear below 1.5mm. We previously presented a fiducial frame for this purpose that attaches to the upper dentition via patient-specific bite blocks and demonstrated a TRE of 0.73mm (±0.23mm) on cadaveric skulls. In that study, TRE measurement depende...
متن کاملQuality Assurance of Multiport Image-Guided Minimally Invasive Surgery at the Lateral Skull Base
For multiport image-guided minimally invasive surgery at the lateral skull base a quality management is necessary to avoid the damage of closely spaced critical neurovascular structures. So far there is no standardized method applicable independently from the surgery. Therefore, we adapt a quality management method, the quality gates (QG), which is well established in, for example, the automoti...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology
دوره 29 8 شماره
صفحات -
تاریخ انتشار 2008