Self-Monitoring of Intraocular Pressure Outside of Normal Office Hours Using Rebound Tonometry: Initial Clinical Experience in Patients With Normal Tension Glaucoma.
نویسندگان
چکیده
PURPOSE To determine the value and accuracy of 24-hour intraocular pressure (IOP) phasing using Icare ONE rebound tonometry (RTONE), in subjects with normal tension glaucoma (NTG). METHODS Eighteen consecutive patients with treated NTG were studied, all subjects had undergone 24-hour IOP phasing during a 1-year period. Each patient had daytime (08:00 to 16:00) IOP phasing with Goldmann applanation tonometer at 2-hourly intervals; at these same time points an IOP reading was also obtained by the patient using RTONE. Self-measured IOPs were then recorded at home using RTONE between 18:00 and 06:00 (at 2-hourly intervals). The frequency with which the phasing results altered clinical management was evaluated. RESULTS The mean peak IOP was significantly higher during nighttime phasing (15.78±4.8 mm Hg) compared with daytime phasing (12.83±2.7 mm Hg, P=0.0018) and clinic IOP measurements (11.8±1.6 mm Hg, P<0.0001). Following IOP phasing a change in management occurred in 10 of 18 patients (56%). In the majority of these patients, a peak IOP was identified during nighttime phasing compared with daytime phasing, this difference was significant (P=0.0090). There were strong correlations between the IOP measurements obtained with Goldmann applanation tonometer and RTONE (Spearman r values >0.60, P<0.001). CONCLUSIONS This study suggests that in patients with NTG with progression that is disproportionate to their clinic IOP measurements, 24-hour phasing can reveal higher IOP spikes than those identified during typical office hours. RTONE is a safe, easy to use, and accurate device for self-monitoring of IOP.
منابع مشابه
Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.
PURPOSE (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer. METHODS A...
متن کاملDiurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial
SIGNIFICANCE This article shows that self-tonometry can provide robust measures of diurnal intraocular pressure (IOP) and also detect changes to IOP in response to treatment within a short period of monitoring. These advances in IOP monitoring may contribute to improved management of glaucoma patients and suspects. PURPOSE The aim of this study was to prospectively investigate the utility of ...
متن کاملThe Application of a Contact Lens Sensor in Detecting 24-Hour Intraocular Pressure-Related Patterns
Glaucoma is one of the leading causes of blindness worldwide. Recent studies suggest that intraocular pressure (IOP) fluctuations, peaks, and rhythm are important factors in disease advancement. Yet, current glaucoma management remains hinged on single IOP measurements during clinic hours. To overcome this limitation, 24-hour IOP monitoring devices have been employed and include self-tonometry,...
متن کاملCorrelation between diurnal variation of intraocular pressure, ocular pulse amplitude and corneal structural properties.
PURPOSE The purpose of this study was to evaluate the fluctuation of intraocular pressure measurements obtained by Goldmann applanation tonometry, dynamic contour tonometry, and corneal compensated non-contact tonometry during office hours in glaucoma and healthy participants. This study also aims at correlating the intraocular pressure fluctuations with fluctuations of corneal hysteresis, cent...
متن کاملCentral corneal thickness measurements in patients with normal tension glaucoma, primary open angle glaucoma, pseudoexfoliation glaucoma, or ocular hypertension.
BACKGROUND/AIMS Recent studies have revealed patients with ocular hypertension to have thicker than normal central corneas and those with normal tension glaucoma to have thinner than normal ones, as determined by ultrasonic pachymetry. Since corneal thickness measurements and applanation tonometric estimates of intraocular pressure (IOP) correlate positively, monitoring of the former parameter ...