Testing a steam-formaldehyde sterilizer for gas penetration efficiency.
نویسندگان
چکیده
A test piece is described for monitoring the performance of low-temperature steam-with-formaldehyde sterilizers. Comparative tests have shown it to be more difficult to penetrate than an arterial catheter when exposed to the same sterilizing conditions. It is permanent and simple to use and maintain. The growth or non-growth of bacterial spores, in the convenient form of spore strips, is used to indicate the efficacy of sterilization.
منابع مشابه
steam-formaldehyde sterilizer for gas penetration efficiency
SYNOPSIS A test piece is described for monitoring the performance of low-temperature steam-withformaldehyde sterilizers. Comparative tests have shown it to be more difficult to penetrate than an arterial catheter when exposed to the same sterilizing conditions. It is permanent and simple to use and maintain. The growth or non-growth of bacterial spores, in the convenient form of spore strips, i...
متن کاملDisinfection of heat-sensitive material by low-temperature steam and formaldehyde.
Steam under subatmospheric pressure at temperatures below 90 degrees C. rapidly killed nonsporing organisms after air had been removed by a high-vacuum pump. Most bacterial spores were killed but small proportions of the populations were very resistant. The destruction of spores was not logarithmic. The addition of formaldehyde vapour to the steam greatly increased its sterilizing power, with d...
متن کاملMeasurement of formaldehyde concentrations in a subatmospheric steam-formaldehyde autoclave.
A method has been developed for measuring formaldehyde concentrations in a subatmospheric steam-formaldehyde autoclave. Data obtained using this method indicate that the concentration of formaldehyde in the chamber atmosphere is not homogeneous and that it decreases rapidly with time. The penetration of formaldehyde vapour into narrow tubes has also been investigated and was shown to be depende...
متن کاملExergy Recovery in Gas Pressure Compression Stations (GPCSs)
The exergy analysis is a proper method for performance evaluation of industrial systems. A generic and detailed analysis of the GPCSs on the second gas pipeline of Iran is made by the means of exergy. The two main improvement measures of fuel pre-heating and steam injection technologies are presented for the current conventional stations. Steady state equations regarding the second law of ther...
متن کاملSynthesis, Characterization, and Efficiency Testing of Ag3PO4/TiO2 Heterogeneous Nano-Photocatalyst in Removing Gaseous Formaldehyde as an Occupational Carcinogen
Introduction: Rapid population growth and industrialization have increased chemical pollutants. Some studies show that employee exposure to formaldehyde in industrial places, hospitals, and laboratory settings is more than the allowed limits. Therefore, it is necessary to implement a proper control system to reduce this exposure. This study aimed to synthesize Ag3PO4/TiO2 nanocomposite, determi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of clinical pathology
دوره 26 9 شماره
صفحات -
تاریخ انتشار 1973