Influence of prosthesis design and implantation technique on implant stresses after cementless revision THR
نویسندگان
چکیده
BACKGROUND Femoral offset influences the forces at the hip and the implant stresses after revision THR. For extended bone defects, these forces may cause considerable bending moments within the implant, possibly leading to implant failure. This study investigates the influences of femoral anteversion and offset on stresses in the Wagner SL revision stem implant under varying extents of bone defect conditions. METHODS Wagner SL revision stems with standard (34 mm) and increased offset (44 mm) were virtually implanted in a model femur with bone defects of variable extent (Paprosky I to IIIb). Variations in surgical technique were simulated by implanting the stems each at 4° or 14° of anteversion. Muscle and joint contact forces were applied to the reconstruction and implant stresses were determined using finite element analyses. RESULTS Whilst increasing the implant's offset by 10 mm led to increased implant stresses (16.7% in peak tensile stresses), altering anteversion played a lesser role (5%). Generally, larger stresses were observed with reduced bone support: implant stresses increased by as much as 59% for a type IIIb defect. With increased offset, the maximum tensile stress was 225 MPa. CONCLUSION Although increased stresses were observed within the stem with larger offset and increased anteversion, these findings indicate that restoration of offset, key to restoring joint function, is unlikely to result in excessive implant stresses under routine activities if appropriate fixation can be achieved.
منابع مشابه
Activity Intensity, Assistive Devices and Joint Replacement Influence Predicted Remodelling in the Proximal Femur
Bone morphology and density changes are commonly observed following joint replacement, and may contribute to the risks of implant loosening and periprosthetic fracture, and reduce the available bone stock for revision surgery. This study was presented in the “Bone and Cartilage Mechanobiology across the scales” WCCM symposium to review the development of remodelling prediction methods and to de...
متن کاملTotal Hip Replacement Revision in a Single Brand Small Cementless Stem – Our Experience after the Findings of the National Joint Registry
Background: Cementless total hip replacement is the common THR performed in England, Wales, Northern Irelandand the Isle of Man. The Corail stem is the most popular cementless implant and has a ODEP 10A rating. Review ofits performance in the registry identified an increase rate of revision amongst the smaller stem sizes. However, claritywas not provided on the explanation for this finding. We ...
متن کاملActivity intensity, assistive devices and joint replacement influence predicted remodelling in the proximal femur.
Bone morphology and density changes are commonly observed following joint replacement, may contribute to the risks of implant loosening and periprosthetic fracture and reduce the available bone stock for revision surgery. This study was presented in the 'Bone and Cartilage Mechanobiology across the scales' WCCM symposium to review the development of remodelling prediction methods and to demonst...
متن کاملIs stem design important in uncemented total hip replacement to decrease subsidence in obese patients?
© Annals of Joint. All rights reserved. Ann Joint 2017;2:50 aoj.amegroups.com There has been a world-wide trend towards cementless total hip replacement (THR) over the last 20 years (1). One of the primary reasons for the introduction of the cementless stem was to improve the outcome in younger, more active patients, particularly males. The New Zealand Joint Registry (NZJR) has shown a revision...
متن کاملProximally coated cementless bipolar hemiarthroplasty in Dorr type C bone.
The current study was conducted to examine the clinical and radiographic outcomes of cementless bipolar arthroplasty in patients who had type C bone at the time of the procedure. A total of 87 patients (105 hips) who had type C femora and had undergone cementless bipolar hemiarthroplasty with a proximally coated cementless prosthesis for the treatment of displaced femoral neck fractures at a si...
متن کامل