Failure of a water-bath design intravenous fluid warmer.
نویسندگان
چکیده
To the Editor, We write to highlight the failure of an intravenous fluid warming device, HOTLINE HL-290 Blood and Fluid Warmer (Smiths Medical International, Kent, UK), which could have resulted in harm to the patient. Fluid warmer failures are becoming more prevalent, as evidenced by the United Kingdom Medicines and Healthcare Products Regulatory Agency (MHRA) issuing a recall of all Level 1 SIMS fluid warmer sets. The fluid warmer in our case is designed differently. Warm water is pumped into a counter-current system around an inner sterile infusion lumen. The warming fluid is kept in a non-sterile water bath, emphasizing the need for extreme care with all water-bath fluid warmers. Our patient received general anesthesia for laparoscopic radical prostatectomy. Apparently, the warming infusion tubing had been tested prior to surgery. However, after 2 L of intravenous fluid had been connected and pressurized, the water bath overflowed. A leak within the coaxial infusion tubing, which allowed mixing of the sterile intravenous fluid and the non-sterile warming fluid, was immediately suspected. The infusion was stopped immediately, and the administration set was disconnected from the patient. An entirely new set was assembled, tested for leaks, and connected, and the device functioned uneventfully for the remainder of the procedure. The patient’s serum electrolyte values remained within normal limits, and the blood cultures were devoid of microbiological growth. Fortunately, the patient completely recovered postoperatively with no signs of sepsis or dilutional electrolyte imbalance. The warming fluid culture revealed heavy growth of pseudomonas and coliform bacteria. Furthermore, cultures from another three water-bath warmers, which were in the theatre suite at the same time, grew diptheroid and fungal species. On subsequent examination of the infusion tubing, a hairline circumferential crack was visible within the wall of the inside lumen. When the water-bath warmer was re-tested, fluid flowed from the outside warming lumen to the inside sterile lumen. We considered this problem to be a manufacturing defect since no direct trauma to the device was observed and the external wall was undamaged. The faulty equipment was returned to the manufacturer for further investigation, and the matter was reported to the MHRA. The manufacturers recommend a number of options for maintaining warming fluid sterility, specifically, if either sterile water or 35% isopropyl alcohol in distilled water is used as the warming fluid, the fluid should be changed and a disinfection protocol followed every 30 days. The alternative is to use a 0.3% hydrogen peroxide solution, which requires the solution to be replaced and the water bath disinfected every 12 months. We use sterile water in our institution, and each hotline is discarded after one use to ensure frequent turnover of the warming fluid. There are important patient safety issues related to this case. It is impossible to know how much of the grossly hypotonic, non-sterile warming fluid entered the patient’s circulation, although the lack of detectable hemolysis, stability of the serum electrolyte measurements, and lack of sepsis postoperatively suggest the amount, if any, was small. Professionals who use this fluid warmer should be alerted to the possibility that warming fluid and intravenous fluid could be mixed. This risk can be eliminated by using the electric hotplate design of fluid warmer, because there is no chance of fluid contamination. If any fluid is observed around the outside of the tubing, then a leak must have occurred. P. A. Clarke, FRCA (&) M. J. Thornton, FRCA Royal United Bath NHS Trust, Bath, Somerset, UK e-mail: [email protected]
منابع مشابه
Effective methods of in-line intravenous fluid warming at low to moderate infusion rates.
Three methods of warming intravenous (IV) fluids were examined. An in-line blood warmer was generally ineffective at flow rates of < 250 mL/hr but did produce temperatures of 30 to 31 degrees C at the catheter when the infusion rate was 500 to 1,000 mL/hr and the tubing was insulated. An in-line hot water bath produced temperatures of > or = 30 degrees C at flow rates of 200 to 1,000 mL/hr with...
متن کاملA Simple Heat Exchanger for Use during Blood Transfusion.
CONSIDERABLE INTEREST has recently been prod to the blood during masswe transfusmn Boyan and Howland of cold bank blood rote an antecubltal veto may produ~ true behind the heart as low as ~7 5 o C The advent c rote the right atrmm has been held responsible foJ anhythmms, particularly ventncular fibrillation dun of blood mtra-operalavely There is no doubt that du human arm acts as an adequate he...
متن کاملAn experimental study of warming intravenous fluid in a cold environment.
OBJECTIVE Numerous studies support the use of warmed intravenous fluids in hypothermic patients. The most effective method to accomplish this goal in a cold prehospital, wilderness, or combat setting is unknown. We evaluated various methods of warming intravenous fluids for a bolus infusion in a cold remote environment. METHODS One liter and 500 mL bags of intravenous fluid at 5 degrees C wer...
متن کاملUltrafiltration in the management of heart failure.
PURPOSE OF REVIEW Congestion causes the majority of hospitalizations for heart failure and contributes to heart failure progression and mortality. Intravenous loop diuretics reduce the signs and symptoms of congestion. Loop diuretics, however, may be associated with increased morbidity and mortality because of deleterious effects on neurohormonal activation, electrolyte balance, and cardiac and...
متن کاملThe use of intravenous therapy.
The aim of this article is to increase your understanding of fluid and electrolyte homeostasis, and provide the rationale for the care and management of patients undergoing intravenous fluid therapy. After reading this article, you should be able to: Describe different body fluid compartments. Define the principles of osmosis, diffusion and filtration in relation to the movement of water and so...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Canadian journal of anaesthesia = Journal canadien d'anesthesie
دوره 56 11 شماره
صفحات -
تاریخ انتشار 2009