Correlation between Glycosyiated Hemoglobin in Specimens Containing Potassium EDTA
نویسندگان
چکیده
We recently started estimating glycosylated hemoglobin (HbA1) by using an agarose electrophoresis system (Corning Medical and Scientific, Palo Alto, CA 94306). It is recommended in the manufacturer’s literature that potassium EDTA-anticoagulated samples should be used. We routinely estimate glucose in fluoride/oxalate anticoagulated samples, so we decided to compare results for HbA1 samples anticoagulated in the two ways. Blood specimens from 80 patients who were attending the diabetic outpatient clinic were collected, part of each in a potassium EDTA-containing tube and another part in a fluoride/oxalatecontaining tube. The samples were then analyzed as detailed by Corning. Each pair of samples was electrophoresed on the same gel to avoid between-batch variation. Comparison of the EDTA-containing samples (x) with the fluoride/oxalate-containing samples (y) gave the following linear regression equation y = 0.97x + 0.59; r = 0.98. A pair difference t-test gave the following result: mean difference 0.26, S.E. ofdifference 0.07, t = 3.670 (p < .001). The difference is statistically significant, but too small to affect clinical interpretation of the results. Consequently, we have started estimating both glucose and HbA1 in fluoride! oxalate-treated samples. This has the obvious advantage to the patient of requiring the withdrawal ofonly one 5mL sample.
منابع مشابه
Correlation between Glycosyiated Hemoglobin in Specimens Containing Potassium EDTA and Fluoride
We recently started estimating glycosylated hemoglobin (HbA1) by using an agarose electrophoresis system (Corning Medical and Scientific, Palo Alto, CA 94306). It is recommended in the manufacturer’s literature that potassium EDTA-anticoagulated samples should be used. We routinely estimate glucose in fluoride/oxalate anticoagulated samples, so we decided to compare results for HbA1 samples ant...
متن کاملScoring System for Detecting Spurious Hemolysis in Anticoagulated Blood Specimens
BACKGROUND The identification of in vitro hemolysis (IVH) using a hematology analyzer is challenging because centrifugation of the specimens cannot be performed for cell counts. In the present study, we aimed to develop a scoring system to help identify the presence of hemolysis in anticoagulated blood specimens. METHODS Thirty-seven potassium EDTA anticoagulated blood specimens were obtained...
متن کاملHeparinized plasma is an unacceptable specimen for ammonia determination.
The glutamate dehydrogenase (GLDH, EC 1.4.1.3) reaction is almost universally used for the determination of ammonia. Some kit manufacturers specify either heparinized or EDTA plasma, whereas others recommend EDTA only, but without saying why. The procedure in Volume 9 of Selected Methods in Clinical Chemisty (1) requires EDTA but does not discuss the consequences of substituting heparin. We hav...
متن کاملCorrelation between Nitric oxide (NO) & Asymmetric dimethylargininie (ADMA) Hemoglobin
73/ Spring summer 2010, Vol.4, No. 1 Medical Laboratory Journal Correlation between Nitric oxide (NO) & Asymmetric dimethylargininie (ADMA) Hemoglobin Concentration in sickle cell patients Abstract Background and objectives: The importance of Nitric oxide (NO) and Asymmetric dimethylargininie (ADMA) in pathophysiology of Sickle cell disease (SCD) is being increasingly clarified. Since very few ...
متن کاملThe effects of different anticoagulants on routine canine plasma biochemistry.
The effects of heparin, ethylenediaminetetraacetic acid (EDTA), sodium citrate and sodium fluoride/potassium oxalate on plasma biochemistry results in dogs were studied and compared with serum. Blood specimens from 10 apparently clinical healthy dogs were collected and placed in different tubes containing each anticoagulant tested. Differences in albumin, acetylcholinesterase, ionized calcium a...