Evaluation of a Method for International Sensitivity ndex Calibration of Two Point-of-Care Prothrombin Time (PT) Monitoring Systems (CoaguChek Mini and TAS PT-NC) with Fresh Plasmas Based on Whole-Blood Equivalent PT
نویسندگان
چکیده
Background: The International Sensitivity Index (ISI) calibration of whole-blood prothrombin time (PT) monitors for point-of-care testing (POCT) described by Tripodi et al. (Thromb Haemost 1993;70:921–4) has been shown to be dependable but is too complex and demanding. The use of plasma would simplify calibration of whole-blood POCT PT monitors, but important differences may exist between the ISI for whole blood and plasma calibrations. Methods: In a 10-center calibration study of two POCT whole-blood monitoring systems (CoaguChek Mini and TAS PT-NC), we characterized the relationship between the log PT for whole blood and fresh plasma with use of single lots of test strips/cards. This relationship (linear) was used to correct the difference between the wholeblood and plasma ISI. The reliability of the correction with different lots of test strips/cards was assessed at three centers. The linear relationship was used to correct the difference in the whole-blood and plasma ISI with four other lots of TAS PT-NC cards and with two additional lots of CoaguChek Mini test strips. Results: The correction decreased the ISI difference from 13.3% to 0.9% for the TAS PT-NC and from 5.7% to 0.6% for the CoaguChek Mini. In assessments at three centers, which included different lots of test strips/ cards, the mean ISI difference was markedly decreased with the TAS PT-NC but not with the CoaguChek Mini, for which the mean ISI difference increased slightly. Conclusions: The proposed correction resolves the discrepancy between whole-blood and fresh plasma ISI calibrations with TAS PT-NC test cards. The CoaguChek Mini systems could be calibrated without this correction. © 2002 American Association for Clinical Chemistry
منابع مشابه
European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors.
AIMS The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma...
متن کاملEuropean concerted action on anticoagulation. Quality assessment of the CoaguChek Mini and TAS PT-NC point-of-care whole-blood prothrombin time monitors.
BACKGROUND International Normalized Ratios (INRs) for prothrombin time obtained with the CoaguChek Mini and TAS (RapidPointCoag) PT-NC systems are markedly different and also differ from the "true" INR. There is therefore a need for local quality assessment (QA) of the two systems. METHODS A set of 60 lyophilized artificially depleted and 60 lyophilized coumarin plasmas were tested at 10 cent...
متن کاملTherapeutic concordance of two portable monitors and two routine automatic oral anticoagulant monitoring systems using as reference the manual prothrombin time technique.
Two models of capillary blood prothrombin time (PT) monitoring systems were evaluated for analytical performance and then compared with two routine PT systems using the reference manual technique and a high-sensitivity thromboplastin. Two sets of 60 and 80 plasmas were analyzed from anticoagulated patients stabilized over 3 months in an INR range 2-3.5 for therapy. Capillary PT determination wa...
متن کاملEuropean concerted action on anticoagulation. Use of plasma samples to derive international sensitivity index for whole-blood prothrombin time monitors.
BACKGROUND To simplify International Sensitivity Index (ISI) calibration, the possibility of substituting fresh plasma for fresh whole-blood samples with point-of-care testing (POCT) whole-blood monitors was investigated in a three-center study of three different POCT systems. METHODS A modified full WHO calibration procedure based on 20 healthy controls and 60 coumarin-treated patients was p...
متن کاملReliability of international normalised ratios from two point of care test systems: comparison with conventional methods.
OBJECTIVE To find out how accurately two point of care test systems--CoaguChek Mini and TAS PT-NC (RapidPointCoag)--display international normalised ratios (INRs). DESIGN Comparison of the INRs from the two systems with a "true" INR on a conventional manual test from the same sample of blood. SETTING 10 European Concerted Action on Anticoagulation centres. PARTICIPANTS 600 patients on lon...